Installation
0112-0127 B
55
A finished installation should look like Figure 2-5.
Figure 2-5:
Restraining cable installation finished
The restraining cable for the fluidics module prevents the
module from pivoting too far, and prevents the accidental
detachment of the serial communication cable. The restraining
cable also holds the module in the open position so that the
instrument can be serviced without first detaching the
communication or power cable.
5.
With the fluidics module tilted back, connect the 15-pin sub-D
electrical connector into the communication port on the far
bottom edge of the fluidics module. It must be aligned properly
to fit.
Figure 2-6:
Fluidics connection labels
6.
Press the connector in firmly.
Note:
The labels on the fluidics connector (
Fluidics Connector
)
and near the communication port (
Connect Fluidics Here
) help
identify these ports.
FlexStation_manual.book Page 55 Tuesday, June 25, 2013 10:22 AM