34
- < STATION APPLICATION > parameter < DRYING > – not editable.
- <REAGENT> parameter <AIR> – not editable.
- <MODE> parameter <+> – modes of heating of air in the drying station.
Pressing the button switches between modes <> (weak) and
<+> (strong).
At the mode <> (weak), the temperature of the airflow is ~ 40ºС (the air
is heated by ~15 – 20 ºС). At the mode <+> (strong), the temperature of the
airflow is ~ 55ºС (the air is heated by ~ 20 – 25 ºС).
The station <FLOW-THROUGH> (station No. 2) has the following configuration
parameters (Fig. 21):
Figure 21. Configuration of station No. 2 <FLOW-THROUGH>.
- < STATION APPLICATION > parameter < FLOW-THROUGH > – not editable.
- <REAGENT> parameter <TAP WATER> – not editable.
- <RESOURCE LIMIT> – not editable (only for station <FLOW-THROUGH>).
- <DELAY> – editable parameter: the duration of a rack’s delay in the upper position
after rising from a trough (0-99 seconds, default value is 10).
Stations No. 3-9 can allocate troughs with reagents or racks, these stations must
have respective application values: <TECHNOLOGICAL> or <PARKING>. All parameters
of stations designated as <TECHNOLOGICAL>, in contrast to <DRYING> and <FLOW-
THROUGH>, are editable (Fig. 22).