Page 27
Section 5.
Troubleshooting
1. Main Power switch does not light - primary pump does not start.
Check main power cable is connected.
Check main fuses (FS1 - located in the power inlet plug on the back
panel).
2. Pump switch lights - primary pump does not start.
Check fuse FS3 - refer to MKS.
Check primary pump temperature. Allow to cool to reset pump
internal thermal trip.
3. Primary pump cuts out.
Check primary pump temperature. Allow to cool to reset pump
internal thermal trip. Check fan filters to ensure sufficient air flow.
4. Pump switch lights - turbo does not start.
Problem with turbo controller/turbo pump - refer to MKS.
Check fuse FS4.
5. Turbo starts - ‘at speed’ indicator does not light.
Poor vacuum due to leak - check all vacuum connections.
Poor vacuum due to primary pump performance - refer to MKS
Problem with turbo controller/turbo pump - refer to MKS.
6. Turbo cuts out.
Poor vacuum due to leak - check all vacuum connections.
Содержание Spectra 138A
Страница 1: ...Instruction Manual MKS Spectra Type 138A Portable Analytical System LP101017 Rev 1 02 April 2002 ...
Страница 9: ...PAS Trolley Manual LP101017 Rev 1 02 Page i Errata and addenda This page is deliberately left blank ...
Страница 10: ...PAS Trolley Manual LP101017 Rev 1 02 Page ii This page is deliberately left blank ...
Страница 17: ...PAS Trolley Manual LP101017 Rev 1 02 Page 9 This page is deliberately left blank ...
Страница 19: ...PAS Trolley Manual LP101017 Rev 1 02 Page 11 ...
Страница 20: ...PAS Trolley Manual LP101017 Rev 1 02 Page 12 This page is deliberately left blank ...
Страница 21: ...PAS Trolley Manual LP101017 Rev 1 02 Page 13 ...
Страница 22: ......
Страница 25: ...PAS Trolley Manual LP101017 Rev 1 02 Page 17 ...
Страница 26: ......
Страница 34: ......
Страница 38: ......
Страница 41: ...PAS Trolley Manual LP101017 Rev 1 02 Page 33 ...
Страница 42: ......
Страница 44: ...PAS Trolley Manual LP101017 Rev 1 02 Page 36 This page is deliberately left blank ...