EtherCAT
®
Communication Protocol Manual
– p/n 20021549-001 Rev. B
17
Index
Sub-
Index
Data Type
Backup(B)/
Save(S)?
Access
PDO
Assignable?
Name
Description
0x18
REAL
B/S
RW
Low Trip Hysteresis
Hysteresis value for Low Trip Point
0x1A
UDINT
B/S
RW
Source Index
Index and sub-index Trip Point Source
—supports only
combination pressure
0xF6401100: Combo Gauge
0x800F
Configuration Trip Point 2
0x01
BOOL
B/S
RW
High Trip Enable
High Trip Point
0: Disable
1: Enable
0x02
BOOL
B/S
RW
Low Trip Enable
Low Trip Point
0: Disable
1: Enable
0x11
REAL
B/S
RW
High Trip Point limit
High limit to trigger trip point condition if Input Value is
above this limit
0x14
REAL
B/S
RW
Low Trip Point Limit
Low limit to trigger trip point condition if Input Value is
below this limit
0x17
REAL
B/S
RW
High Trip Hysteresis
Hysteresis value for High Trip Point
0x18
REAL
B/S
RW
Low Trip Hysteresis
Hysteresis value for Low Trip Point
0x1A
UDINT
B/S
RW
Source Index
Supports only Combination Pressure as the Trip Point
Source.
0xF6401100: Combo Gauge
0x9003
RO
Information Heat Transfer
See Sensor Warnings and Errors
0x01
UINT
RO
Sensor Warnings
0x02
UINT
RO
Sensor Errors
0x900E
Information Trip Point 1
0x01
REAL
RO
Input Value Trip Point 1
Trip Point input value as referenced by 800E:1A
0x900F
Information Trip Point 2
0x01
REAL
RO
Input Value Trip Point 2
Trip Point input value as referenced by 800F:1A
0x9015
Information Hot Cathode Ion
Gauge
0x01
UINT
RO
Sensor Warnings
See Sensor Warnings and Errors
0x02
UINT
RO
Sensor Errors
See Sensor Warnings and Errors
0x06
UINT
RO
Degas Time on remaining
Remaining time for the current degas on cycle in seconds