5 DeviceNet Operation
30
Series 275 Convectron Module with DeviceNet
Instruction Manual - 275502
Polled Data
A network POLL will return data as A/D counts. This response can be converted to pressure as
indicated in Table 5-7. The maximum value is 3FFF or 2^14 counts.
Table 5-7
Polled Data to Pressure Conversion
Pressure
(Torr)
A/D Counts
(decimal)
A/D Counts
(Hex)
0.0
0
0
0.0001
14
E
0.0002
30
1E
0.0005
78
4E
0.001
157
9D
0.002
311
137
0.005
749
2ED
0.01
1424
590
0.0193
2500
9C4
0.02
2515
9D3
0.05
2888
B48
0.1
3349
D15
0.2
4002
FA2
0.5
5251
1483
1
6523
197B
2
7997
1F3D
5
9958
26E6
10
11210
2BCA
20
12090
2F3A
50
12733
31BD
100
12963
32A3
200
13137
3351
300
13352
3428
400
13627
353B
500
13874
3632
600
14089
3709
700
14266
37BA
760
14360
3818
800
14416
3850
900
14547
38D3
1000
14653
393D
Содержание 275420-1-GQ-T-00011
Страница 2: ......
Страница 16: ...2 Introduction Specifications 16 Series 275 Convectron Module with DeviceNet Instruction Manual 275502...
Страница 26: ...4 Operation 26 Series 275 Convectron Module with DeviceNet Instruction Manual 275502 Notes...
Страница 31: ...5 DeviceNet Operation Series 275 Convectron Module with DeviceNet Instruction Manual 275502 31...
Страница 32: ...5 DeviceNet Operation 32 Series 275 Convectron Module with DeviceNet Instruction Manual 275502...
Страница 50: ......
Страница 51: ......