1579A/579A
Chapter 6: Operation (Profibus Units)
47
Send Data
Name: SET_MFC
Type:
Cyclic Write (small & full)
Size: 6
Description:
Analog Output Transducer Block, Small & Full Setup
Parameters:
Name
Add.
Type
Comment
VALVE_OVERRIDE
1:0
uint:2
NORMAL, FLOW_OFF, PURGE
AUTOZERO
1:2
uint:1
0 to 1 transition activates zeroing if
(VALVE_OVERRIDE==FLOW_OFF &&
FLOW_SETPOINT < 5%FS)
REPORT_DIAG
1:3
uint:3
transition to a new value, triggers the device to
send a new actual diagnosis:
0 =
no
diagnosis
1
= diagnosis of small functionality
2
= diagnosis of full functionality
3
= report selected gas table
4..7 = reserved
WINK_STATUS
1:6
uint:1
0 to 1 transition sets the LED to blink red/green
for 3 seconds
ENABLE_TOTALIZER
1:7
uint:1
0 = disabled, 1 = enabled
RESET_TOTALIZER
0:0
uint:1
0 to 1 transition resets totalizer to zero
RESET_STATUS
0:1
uint:1
0 to 1 transition resets error status bits
SELECT_GAS_TABLE
0:2 uint:4 0..14; 15 = default gas table is used
EN_GAS_CORRECTION 0:6
uint:1
0 = disabled, 1 = enabled
Reserved
0:7
uint:1
FLOW_SETPOINT
2:0
long
in [FLOW_UNIT] in 10E-4 steps
valve switched off if setpoint < 1%
valve switched on if setpoint > 2%
If the device is in the analog mode (i.e. set point is taken from the analog interface) some functions
are not active:
1. VALVE_OVERRIDE is not active
2. EN_GAS_CORRECTION is not active
3. SELECT_GAS_TABLE is not active and the default table will be used
Содержание 1579A
Страница 8: ...Table of Contents 1579A 579A viii This page intentionally left blank...
Страница 12: ...Safety Information 1579A 579A 12 This page intentionally left blank...
Страница 16: ...Chapter 1 General Information 1579A 579A 16 This page intentionally left blank...
Страница 20: ...Chapter 2 Shipment 1579A 579A 20 This page intentionally left blank...
Страница 28: ...Chapter 3 Connectors and Control Elements 1579A 579A 28 This page intentionally left blank...
Страница 38: ...Chapter 5 Operation Analog Units 1579A 579A 38 This page intentionally left blank...
Страница 70: ...Chapter 10 Maintenance 1579A 579A 70 This page intentionally left blank...
Страница 74: ...Appendix A Product Specifications 1579A 579A 74 This page intentionally left blank...
Страница 82: ...Appendix C Gas Correction Factors 1579A 579A 82 This page intentionally left blank...
Страница 84: ...Appendix C Gas Correction Factors 1579A 579A 84 This page intentionally left blank...
Страница 86: ...Next page Health and Safety Form Contact MKS at the web or at any MKS location if the form is missing...