Chapter 3: Connectors and Control Elements
1579A/579A
22
ZERO
SETUP
RESET
AUTOZERO
Figure 4: 1579A/B / 579A/B Gas Inlet Side
Potentiometer ZERO
Use this 25 turn potentiometer to adjust the signal output to zero.
Refer to
Chapter 5, Zero Adjust
.
AUTOZERO
Instead of potentiometer ZERO you can use this push button to zero the signal output.
Refer to
Chapter 5, Zero Adjust
.
RESET Switch
After pressing the RESET key the unit performs a self check cycle and a reset. The same
procedure is done after power on. During the cycle the two color LED, labeled STS, changes its
color a few times between red and green.
Finally the LED changes to green, indicating that self test cycle and reset have been completed
successfully.
Connector SETUP
This access allows to modify the setup via special adapter and WINDOWS software. These tools
may be ordered as a kit, ordering code DMC-SUPPORT-ANA. The adapter only is available as
ADAPTER-DMC-ANA.
Note:
Warranty is void when different connector or cable is used!
Содержание 1579A
Страница 8: ...Table of Contents 1579A 579A viii This page intentionally left blank...
Страница 12: ...Safety Information 1579A 579A 12 This page intentionally left blank...
Страница 16: ...Chapter 1 General Information 1579A 579A 16 This page intentionally left blank...
Страница 20: ...Chapter 2 Shipment 1579A 579A 20 This page intentionally left blank...
Страница 28: ...Chapter 3 Connectors and Control Elements 1579A 579A 28 This page intentionally left blank...
Страница 38: ...Chapter 5 Operation Analog Units 1579A 579A 38 This page intentionally left blank...
Страница 70: ...Chapter 10 Maintenance 1579A 579A 70 This page intentionally left blank...
Страница 74: ...Appendix A Product Specifications 1579A 579A 74 This page intentionally left blank...
Страница 82: ...Appendix C Gas Correction Factors 1579A 579A 82 This page intentionally left blank...
Страница 84: ...Appendix C Gas Correction Factors 1579A 579A 84 This page intentionally left blank...
Страница 86: ...Next page Health and Safety Form Contact MKS at the web or at any MKS location if the form is missing...