Chapter 3: Connectors and Control Elements
1579A/579A
26
Electrical Connectors
1579A/B / 579A/B, Profibus Units, Connector ANALOG
(9 pol. Sub-D, male)
Pin
± 15 V Supply
24 V Supply
Note
1
Valve Open / Valve Close
1,2
Valve Open / Valve Close
2,3
2
Flow Signal
(0 to +5 V)
Analog operation not
possible
Referenced to pin 7 and 8
3
+ 15 V Supply
+ 24 V Supply
4
±
15 V Common
reserved
5
- 15 V Supply
24 V Common
6
Setpoint Input
2
Analog operation not
possible
Referenced to pin 7 and 8
7
Signal Common
-
8
Signal Common
-
9
Reserved Reserved
Figure 9: 9-pol. Connector
For the use of the Profibus version in analog mode is a +/- 15 VDC power supply required. For the
use in digital mode a +/- 15 VDC or a 24 VDC power supply is sufficient. This connector does not
provide a particular pin for connection of the shield to the case. Connect the cable shield to the
connector housing for grounding.
The cable must be equipped with a ferrite (Würth 742 712 21).
_______________
1
To close the valve connect pin 1 to pin 5 (-15 V);
To open the valve apply + 5 V to pin 1 (referenced to pin 7 or 8).
2
Not applicable with mass flow meter 579A/B
3
To close the valve connect pin 1 to pin 4;
To open the valve apply + 5 V to pin 1 (referenced to pin 7 or 8).
Содержание 1579A
Страница 8: ...Table of Contents 1579A 579A viii This page intentionally left blank...
Страница 12: ...Safety Information 1579A 579A 12 This page intentionally left blank...
Страница 16: ...Chapter 1 General Information 1579A 579A 16 This page intentionally left blank...
Страница 20: ...Chapter 2 Shipment 1579A 579A 20 This page intentionally left blank...
Страница 28: ...Chapter 3 Connectors and Control Elements 1579A 579A 28 This page intentionally left blank...
Страница 38: ...Chapter 5 Operation Analog Units 1579A 579A 38 This page intentionally left blank...
Страница 70: ...Chapter 10 Maintenance 1579A 579A 70 This page intentionally left blank...
Страница 74: ...Appendix A Product Specifications 1579A 579A 74 This page intentionally left blank...
Страница 82: ...Appendix C Gas Correction Factors 1579A 579A 82 This page intentionally left blank...
Страница 84: ...Appendix C Gas Correction Factors 1579A 579A 84 This page intentionally left blank...
Страница 86: ...Next page Health and Safety Form Contact MKS at the web or at any MKS location if the form is missing...