22
4
Operatio
n
Using with the infrared mode
CAUTION
The operating table does not function if the IR code
displayed on the touch panel does not match with
that on the control unit.
When there are plural operating tables in one room,
using the same IR code for them possibly causes
cross talk resulting in malfunction. In this case, set
the IR code for each operating table to a different
number from the others.
1.
Open the top screen of the touch panel.
2.
Tap "WIRELESS SETTING".
The "WIRELESS SETTING" screen appears.
3.
Tap "IR".
4.
Disconnect the control unit.
5.
Press and hold
for 10 seconds.
An IR code appears on the monitor screen of the control unit.
NOTE
For how to disconnect the
control unit, refer to Page
21