Cause

Solution

Symptom 3: The Counter executes Zero setting unexpectedly. 

A power interruption (sudden 

drop in power voltage) has 

occurred.

Check the power supply. For details 

about how to cancel the error, see 

 

[------] in the List of Error Codes. 

 

"6.2 List of Error Codes" (page 8)

Symptom 4: BCD output is not working. 

BCD output is not enabled.

Enable BCD output. (Set the value of 

parameter number 20 to 1.)

Output logic is not correct.

Select the output logic that is 

appropriate for your application by 

setting parameter number 24.

Command mode is active.

Enable interval mode. (Set the value 

of parameter number 22 to 1.)

Symptom 5: Cannot modify the parameter settings. 

"Set parameters" (set value 1) 

has not been set for Parameter 

mode.

Set the value for parameter number 

00 to 1. For details, see 

"3.1 

Procedure for Setting Parameters" 

(page 3).

6.2 

List of Error Codes

Display NOM 

signal

BCD 

output

Cause

Solution/ 

Error cancellation 

method

Err 10

H

FFFF10

Abnormal power 

voltage

Connect to the specified 

power.
Automatic cancellation

[------]

H

FFFF15

In stand-by state 

after power-

on or a power 

interruption

Re-check the power if a 

power interruption has 

occurred.

 

y

Press [P.SET].

 

y

Input an external HOLD 

signal.

Err 20

H

FFFF20

Excess speed

Revise the measurement 

conditions.

 

y

Press [P.SET].

 

y

Input an external HOLD 

signal.

Err 30

H

FFFF30 Counter value is 

8 digits or more

Modify the Preset value.

 

y

Press [P.SET].

 

y

Input an external HOLD 

signal.

Err 40

H

FFFF40 Linear Gage 

malfunction or 

excess speed

 

y

Check the Linear Gage 

connection.

 

y

Revise the 

measurement 

conditions.

 

y

Press [P.SET].

 

y

Input an external HOLD 

signal.

F*****

L

F*****

Counter value is 

6 digits or more

Return the counter value to 

within 6 digits.

All 

decimal 

points 

blinking

L

Counter 

value 

status

The origin is not 

detected yet 

(EG-101Z only)

Pass the spindle of the 

Linear Gage through the 

origin.
Automatic cancellation

Err 90

L

Counter 

value 

status

Tolerance value 

setting error

Input the tolerance value 

again.

Press [P.SET].

Err 95

L

Counter 

value 

status

Key protect

Cancel the key protection 

parameter. (Set parameter 

number 35 to 0.)

Automatic cancellation

Tips

• 

If the NOM signal is "H": When an error occurs in 3-step or 5-step tolerance 

judgment output (the value for parameter number 20 is 0), both pin numbers 3 

(L1) and 7 (L5) will be "L".

• 

If an error occurs while you are setting parameters, the Preset value, or the 

tolerance value, the error will be displayed after you return to the counter state. 

However, the error signal will be output immediately to any external output.

•  Errors can also be canceled with an external PSET signal.
• 

For EG-101D, all decimal points blink for about 8 seconds when an error is 

canceled.

Specifications

 

„

Basic specifications

Code No.

542-015

542-017

542-016

Model No.

EG-101P

EG-101Z

EG-101D

Number of display axis

1 axis

Display

Minus (–) sign and 6 numeric digits (green LED)

Minimum reading 

(Count display range)

Selected by parameter

Set 

automatically 

according to 

the gage

0.01 (

±9999.99) mm  

0.005 (

±

999.995) mm 

 

0.001 (

±999.999) mm  

0.0005 (

±

99.9995) mm 

 

0.0001 (

±99.9999) mm

Maximum input 

frequency

1.25 MHz 

 

(2-phase square wave)

Maximum count speed

5 MHz

Power source voltage

DC +12 V to 24 V

Power consumption

Max. 6 W (Max. 500 mA)

Operating temperature 

(humidity) range

0 °C to 40 °C

 (20 

%

 RH to 80 % RH, without 

condensation)

Storage temperature 

(humidity) range

-10 

°C

 to 50 

°C

 (20 % RH to 80 % RH, without 

condensation)

External dimensions

96 (W) ×

 48 (H) 

×

 155 (D) mm

Mass

Approx. 400 g

CE marking

EMC directive EN61326-1

 

 

Immunity test requirement: Clause 6.2 Table 2 

 

 

Emission limit: Class B 

 

RoHS directive EN50581

Interface

I/O

 

„

External dimensions drawing (for all models)

154.6

48

 

96 

44

.4

91.4

Unit: mm

 

„

Options

Part No.

Name

02ADB440

I/O output connector (with cover)

02ADD930

Terminal strip connecting cable*

1

357651

AC adapter

02ZAA000

AC cable*

1

*

1

 Required if using the AC adapter.

Mitutoyo Corporation

20-1, Sakado, 1-Chome, Takatsu-ku, Kawasaki-shi, 

Kanagawa 213-8533, Japan

Содержание EG-101P

Страница 1: ...nted Precautions for Use Do not remove the cover or disassemble the product Otherwise you may be subject to electric shock and there is a risk of breakage or fire due to a short circuit caused by meta...

Страница 2: ...ear Gage OUTPUT connector I O connector For connecting an I O connecting cable Grounding terminal For connecting a grounding wire Power terminal block For connecting the Terminal strip connecting cabl...

Страница 3: ...number to 12 The current value of parameter number 12 will blink Parameter number 12 sets the resolution Parameter number Set value 5 Repeatedly press P SET to set the set value to 1 resolution 5 m T...

Страница 4: ...p For EG 101P this is not available 0 Disabled 1 Enabled Parameter number Setting item Description the values in bold indicate the default value 43 Origin initialization when the power is turned on EG...

Страница 5: ...tolerance judgment function is disabled when BANK0 is selected 4 4 Tolerance Value Setting There are 2 settings for the tolerance value 3 step and 5 step Tips Set the value of parameter number 21 to...

Страница 6: ...st significant digit stops blinking 8 Press and hold Fn and then press P SET The Counter will return to the Counter display 5 External Input Output Function This product has an I O connector that enab...

Страница 7: ...e BANK specification MODE BANK SET1 2 3 Min 10 ms Min 10 ms Min 5 ms Max 10 ms DATA BANK pin number 36 BANK switchover MODE pin number 29 Peak switchover SET2 SET1 BANK0 H H BANK1 H L BANK2 L H BANK3...

Страница 8: ...the origin Automatic cancellation Err 90 L Counter value status Tolerance value setting error Input the tolerance value again Press P SET Err 95 L Counter value status Key protect Cancel the key prot...

Отзывы: