5
Printed in Japan
Date of publication: February 1, 2019
7
Press [Fn].
» The Counter will return to the Counter display.
8
Press [P.SET].
» The current value will be changed to the Preset value that was set.
4.2
Peak Mode Setting
The maximum value, the minimum value, and TIR value are constantly
calculated in the Counter. By switching the mode, you can display the
counter value according to the intended application.
Mode
Description
Normal
measurement
Counts the movement (displacement) of the contact
point of the Linear Gage, and then displays the
value successively.
Max. hold
measurement
Displays the maximum value (MAX) measured
during the measurement. The display will not
change until a new maximum value is measured.
Min. hold
measurement
Displays the minimum value (MIN) measured during
the measurement. The display will not change until
a new minimum value is measured.
TIR measurement
Displays TIR value during the measurement = TIR
(maximum value - minimum value). The display will
not change until either a new maximum value or
minimum value is measured.
Procedure for switching Peak mode
1
Repeatedly press [MODE] until the desired mode is displayed.
» The mode will switch as follows:
Normal
measurement
Max. hold
measurement
Min. hold
measurement
TIR
measurement
MAX
MIN
TIR
MAX
MIN
TIR
MAX
MIN
TIR
MAX
MIN
TIR
Procedure for clearing the peak value
1
Press [MODE].
» Peak mode will be set.
2
Press [P.SET].
»
The peak value will be cleared (MAX = MIN = current value, TIR = 0).
4.3
Switching the Tolerance Bank
When setting 3-step or 5-step tolerance values, 3 sets of tolerance value
settings can be saved in internal counter memory called a BANK. You can
recall the saved tolerance value settings by switching the BANK.
Tips
•
For details about tolerance value settings, see
•
The BANK can also be switched by an external signal.
This section explains how to switch the BANK.
1
Press [BANK].
»
The BANK number will be switched.
»
The BANK indicator will switch according to the selected BANK number as
shown in the figure below.
BANK0
BANK1
BANK2
BANK3
BANK
BANK
BANK
BANK
Tips
•
The BANK indicator will switch in order from BANK0 through BANK3 each time
you press [BANK].
•
While [BANK] is pressed down, the currently selected BANK number is
displayed. When the key is released, the Counter will return to the Counter
display.
•
The tolerance judgment function is disabled when BANK0 is selected.
4.4
Tolerance Value Setting
There are 2 settings for the tolerance value: 3-step and 5-step.
Tips
•
Set the value of parameter number 21 to 0 (3-step tolerance) or 1 (5-step
tolerance) in advance.
•
For details about I/O output, see
"5 External Input/Output Function" (page 6).
3-step tolerance value setting (3-step tolerance zone selection)
With S1 and S4 set as the tolerance values, the 3-step tolerance judgment
will be performed as follows:
Judgment conditions
BANK indicator
I/O output
(PIN number)
Measurement result < S1
Amber indicator on
L1 (3)
S1 ≤ measurement result ≤ S4
Green indicator on
L3 (5)
S4 < measurement result
Red indicator on
L5 (7)
This section explains how to set the 3-step tolerance value.
1
Press [BANK] to select the BANK number that you want to set.
»
The BANK indicator corresponding to the selected BANK number will light.
Tips
Select a BANK number from BANK1 through 3. Tolerance values cannot be set if
BANK0 is selected.
2
Press [Fn] to switch to setup mode.
»
The BANK indicator will light in amber. (Tolerance value S1 will be selected.)
3
Press [MODE].
» The input digit will shift to the right. (The
currently selected digit will blink.)
4
Press [P.SET].
»
The tolerance value will be modified.
Tips
The ±
sign is also set at the most significant digit. To set the tolerance value to a
negative value, repeatedly press [P.SET] until the Sign indicator lights.
5
Repeat step
3
and step
4
until the least significant digit has
been set.
»
The least significant digit will blink.
6
Press [MODE].
»
Tolerance value S1 will be applied. (The
least significant digit stops blinking.)
7
Press [Fn].
»
The BANK indicator will light in red. (Tolerance value S4 will be selected.)
8
Set the tolerance value S4 in the same steps as in
3
to
5
.
9
Press [MODE].
»
Tolerance value S4 will be applied. (The least significant digit stops blinking.)
10
Press [Fn].
» The Counter will return to the Counter display.
Tips
An error will occur unless S1 ≤ S4. Press [P.SET] to redo the input from S1.
5-step tolerance value setting (5-step tolerance zone selection)
With S1 to S4 set as the tolerance values, the 5-step tolerance judgment will
be performed as follows:
Judgment conditions
BANK indicator
I/O output
(PIN number)
Measurement result < S1
Amber indicator on
L1 (3)
S1 ≤ measurement result < S2
Amber indicator blinks
L2 (4)
S2 ≤ measurement result ≤ S3
Green indicator on
L3 (5)
S3 < measurement result ≤ S4
Red indicator blinks
L4 (6)
S4 < measurement result
Red indicator on
L5 (7)
This section explains how to set the 5-step tolerance value.