Press [Fn].

 

» The Counter will return to the Counter display.

Press [P.SET].

 

» The current value will be changed to the Preset value that was set.

4.2 

Peak Mode Setting

The maximum value, the minimum value, and TIR value are constantly 

calculated in the Counter. By switching the mode, you can display the 

counter value according to the intended application.

Mode

Description

Normal 

measurement

Counts the movement (displacement) of the contact 

point of the Linear Gage, and then displays the 

value successively.

Max. hold 

measurement

Displays the maximum value (MAX) measured 

during the measurement. The display will not 

change until a new maximum value is measured.

Min. hold 

measurement

Displays the minimum value (MIN) measured during 

the measurement. The display will not change until 

a new minimum value is measured.

TIR measurement

Displays TIR value during the measurement = TIR 

(maximum value - minimum value). The display will 

not change until either a new maximum value or 

minimum value is measured.

 

„

Procedure for switching Peak mode

Repeatedly press [MODE] until the desired mode is displayed.

 

» The mode will switch as follows: 

Normal 

measurement

Max. hold 

measurement

Min. hold 

measurement

TIR 

measurement

MAX

MIN

TIR

MAX

MIN

TIR

MAX

MIN

TIR

MAX

MIN

TIR

 

„

Procedure for clearing the peak value

Press [MODE].

 

» Peak mode will be set.

Press [P.SET].

 

»

The peak value will be cleared (MAX = MIN = current value, TIR = 0).

4.3 

Switching the Tolerance Bank

When setting 3-step or 5-step tolerance values, 3 sets of tolerance value 

settings can be saved in internal counter memory called a BANK. You can 

recall the saved tolerance value settings by switching the BANK.

Tips

• 

For details about tolerance value settings, see 

"4.4 Tolerance Value 

Setting" (page 5).

• 

The BANK can also be switched by an external signal.

This section explains how to switch the BANK.

Press [BANK].

 

»

The BANK number will be switched.

 

»

The BANK indicator will switch according to the selected BANK number as 

shown in the figure below.

BANK0

BANK1

BANK2

BANK3

BANK

BANK

BANK

BANK

Tips

• 

The BANK indicator will switch in order from BANK0 through BANK3 each time 

you press [BANK].

• 

While [BANK] is pressed down, the currently selected BANK number is 

displayed. When the key is released, the Counter will return to the Counter 

display.

• 

The tolerance judgment function is disabled when BANK0 is selected.

4.4 

Tolerance Value Setting

There are 2 settings for the tolerance value: 3-step and 5-step.

Tips

• 

Set the value of parameter number 21 to 0 (3-step tolerance) or 1 (5-step 

tolerance) in advance.

• 

For details about I/O output, see 

"5 External Input/Output Function" (page 6).

 

„

3-step tolerance value setting (3-step tolerance zone selection)

With S1 and S4 set as the tolerance values, the 3-step tolerance judgment 

will be performed as follows:

Judgment conditions

BANK indicator

I/O output  

(PIN number)

Measurement result < S1

Amber indicator on

L1 (3)

S1 ≤ measurement result ≤ S4

Green indicator on

L3 (5)

S4 < measurement result

Red indicator on

L5 (7)

This section explains how to set the 3-step tolerance value.

Press [BANK] to select the BANK number that you want to set.

 

»

The BANK indicator corresponding to the selected BANK number will light.

Tips

Select a BANK number from BANK1 through 3. Tolerance values cannot be set if 

BANK0 is selected.

Press [Fn] to switch to setup mode.

 

»

The BANK indicator will light in amber. (Tolerance value S1 will be selected.)

Press [MODE].

 

» The input digit will shift to the right. (The 

currently selected digit will blink.)

Press [P.SET].

 

»

The tolerance value will be modified.

Tips

The ±

 sign is also set at the most significant digit. To set the tolerance value to a 

negative value, repeatedly press [P.SET] until the Sign indicator lights.

Repeat step 

3

 and step 

4

 until the least significant digit has 

been set.

 

»

The least significant digit will blink.

Press [MODE].

 

»

Tolerance value S1 will be applied. (The 

least significant digit stops blinking.)

Press [Fn].

 

»

The BANK indicator will light in red. (Tolerance value S4 will be selected.)

Set the tolerance value S4 in the same steps as in 

3

 to 

5

.

Press [MODE].

 

»

Tolerance value S4 will be applied. (The least significant digit stops blinking.)

10 

Press [Fn].

 

» The Counter will return to the Counter display.

Tips

An error will occur unless S1 ≤ S4. Press [P.SET] to redo the input from S1.

 

„

5-step tolerance value setting (5-step tolerance zone selection)

With S1 to S4 set as the tolerance values, the 5-step tolerance judgment will 

be performed as follows:

Judgment conditions

BANK indicator

I/O output  

(PIN number)

Measurement result < S1

Amber indicator on

L1 (3)

S1 ≤ measurement result < S2

Amber indicator blinks

L2 (4)

S2 ≤ measurement result ≤ S3

Green indicator on

L3 (5)

S3 < measurement result ≤ S4

Red indicator blinks

L4 (6)

S4 < measurement result

Red indicator on

L5 (7)

This section explains how to set the 5-step tolerance value.

Содержание EG-101P

Страница 1: ...nted Precautions for Use Do not remove the cover or disassemble the product Otherwise you may be subject to electric shock and there is a risk of breakage or fire due to a short circuit caused by meta...

Страница 2: ...ear Gage OUTPUT connector I O connector For connecting an I O connecting cable Grounding terminal For connecting a grounding wire Power terminal block For connecting the Terminal strip connecting cabl...

Страница 3: ...number to 12 The current value of parameter number 12 will blink Parameter number 12 sets the resolution Parameter number Set value 5 Repeatedly press P SET to set the set value to 1 resolution 5 m T...

Страница 4: ...p For EG 101P this is not available 0 Disabled 1 Enabled Parameter number Setting item Description the values in bold indicate the default value 43 Origin initialization when the power is turned on EG...

Страница 5: ...tolerance judgment function is disabled when BANK0 is selected 4 4 Tolerance Value Setting There are 2 settings for the tolerance value 3 step and 5 step Tips Set the value of parameter number 21 to...

Страница 6: ...st significant digit stops blinking 8 Press and hold Fn and then press P SET The Counter will return to the Counter display 5 External Input Output Function This product has an I O connector that enab...

Страница 7: ...e BANK specification MODE BANK SET1 2 3 Min 10 ms Min 10 ms Min 5 ms Max 10 ms DATA BANK pin number 36 BANK switchover MODE pin number 29 Peak switchover SET2 SET1 BANK0 H H BANK1 H L BANK2 L H BANK3...

Страница 8: ...the origin Automatic cancellation Err 90 L Counter value status Tolerance value setting error Input the tolerance value again Press P SET Err 95 L Counter value status Key protect Cancel the key prot...

Отзывы: