   The same function as in tolerance judgment output mode.

*

2

   "L" when output data is fixed.

*

3

 

   During input, the output of pin numbers 3 through 26 and pin number 31 is "H".

 

„

Output circuit

Transistor is on when the output is "L" (open collector).

COM

0.01 µF

Counter

TD62583 equivalent

Output

External 
equipment 

Reference 
circuit

Output withstand voltage: Max. 24 V
Output current: Max. 10 mA
Output saturation voltage: Max. 0.7 V

Surge-current-absorbing diode

60 V 100 mA or more

• 

When using relays, incorporate a surge-current-absorbing diode or a protective 

circuit. If no protection is incorporated, the IC in the Counter may be damaged.

•  The output current when the tolerance judgment result is output is  

20 mA at maximum.

 

„

Input circuit

Input is valid when the input voltage is "L".

+5 V

5 kΩ

5 kΩ

0.01 µF

Input current: Max. 1 mA
Input voltage:   H = 4 to 24 V

 

L = Max. 1 V

Counter

External 
equipment

Use open-collector 
output or relay 

output, etc.

Reference 
circuit

5.2 

Timing Chart

 

z

Tolerance judgment result output

Max. 10 ms 

Max. 10 ms 

+NG

-NG

Upper specification limit

Lower specification limit

Counter data

Tips

• 

After acquiring the counter data, there is a maximum 10 ms delay before the 

tolerance judgment result is output.

• 

For EG-101D, the length of time until the tolerance judgment result is output 

after the counter data enters in the Specification range depends on the 

connected equipment, such as the Linear Gage.

 

z

Preset, Peak clear

PSET

Min. 10 ms 

Min. 

10 ms  

Max. 10 ms 

BCD data

 

z

Peak mode/BANK specification

MODE

 

BANK

SET1, 2, 3

Min. 10 ms

Min. 

10 ms  

Min. 

5 ms  

Max. 10 ms 

DATA

BANK (pin number 36): 

BANK switchover

MODE (pin number 29): 

Peak switchover

SET2

SET1

BANK0

H

H

BANK1

H

L

BANK2

L

H

BANK3

L

L

SET2

SET1

NOMAL

H

H

MAX

H

L

MIN

L

H

TIR

L

L

 

z

HOLD

Error cancellation

(For EG-101Z: origin detection wait state)

HOLD

Min. 10 ms 

Min. 

10 ms  

Max. 15 ms 

BCD data

Tips

During HOLD input, the UNIT indicator will blink.

 

z

Interval mode

Continuously outputs data using the Counter's internal timing.

*1 Depends on the setting of Parameter number 23

READY

(T1) Min. 5/15/20/40 ms*1 

(T2) Min. 2/5/8/14 ms*1

 

BCD data

 

z

Command mode

Outputs data using sync control via HOLD and READY.

READY

HOLD

Min. 10 ms 

Max. 15 ms 

(T2) 

Data latch max. 10 µs (EG-P/Z)

BCD data

 

z

INH input

BCD data output is switched off during INH input.

INH

Min. 10 ms

 

Max. 10 ms

 

Max. 10 ms

READY

BCD data

Troubleshooting

6.1 

Troubleshooting

When the Counter does not operate as expected, refer to the cause of the 

trouble and the solutions shown below:

Cause

Solution

Symptom 1: The counter value is incorrect (not counting).

Parameters are not correctly 

set for the type of the Linear 

Gage, etc.

Set correct parameters. For details, 

see 

"3.2 Basic Parameters" (page 

3).

Peak mode (MAX or MIN is lit) 

is active.

Cancel Peak mode. For details, see 

"4.2 Peak Mode Setting" (page 5).

The HOLD signal (UNIT is 

blinking) is being input.

Check the external input.

Calculation with a constant 

function is set.

Cancel calculation with a constant 

function. (Set parameter number 16 to 

0.)

Symptom 2: Cannot execute Zero setting. 

Peak mode is active.

Cancel Peak mode. For details, see 

"4.2 Peak Mode Setting" (page 5).

The Preset value is a value 

other than 0.

Set the Preset value to 0. For details, 

see 

"4.1 Preset" (page 4).

Содержание EG-101P

Страница 1: ...nted Precautions for Use Do not remove the cover or disassemble the product Otherwise you may be subject to electric shock and there is a risk of breakage or fire due to a short circuit caused by meta...

Страница 2: ...ear Gage OUTPUT connector I O connector For connecting an I O connecting cable Grounding terminal For connecting a grounding wire Power terminal block For connecting the Terminal strip connecting cabl...

Страница 3: ...number to 12 The current value of parameter number 12 will blink Parameter number 12 sets the resolution Parameter number Set value 5 Repeatedly press P SET to set the set value to 1 resolution 5 m T...

Страница 4: ...p For EG 101P this is not available 0 Disabled 1 Enabled Parameter number Setting item Description the values in bold indicate the default value 43 Origin initialization when the power is turned on EG...

Страница 5: ...tolerance judgment function is disabled when BANK0 is selected 4 4 Tolerance Value Setting There are 2 settings for the tolerance value 3 step and 5 step Tips Set the value of parameter number 21 to...

Страница 6: ...st significant digit stops blinking 8 Press and hold Fn and then press P SET The Counter will return to the Counter display 5 External Input Output Function This product has an I O connector that enab...

Страница 7: ...e BANK specification MODE BANK SET1 2 3 Min 10 ms Min 10 ms Min 5 ms Max 10 ms DATA BANK pin number 36 BANK switchover MODE pin number 29 Peak switchover SET2 SET1 BANK0 H H BANK1 H L BANK2 L H BANK3...

Страница 8: ...the origin Automatic cancellation Err 90 L Counter value status Tolerance value setting error Input the tolerance value again Press P SET Err 95 L Counter value status Key protect Cancel the key prot...

Отзывы: