[Parameter number]
/Setting item
Description
(the values in bold indicate the default value)
[10]
Parameter
initialization*
1
If you set the value of this parameter to 1, the
set values for all parameters, except for the
unit setting, can be reset to their default values
(initialized).
Once this setting has been enabled, this parameter
is reset so its set value is 0 (do not initialize).
0: Do not initialize
1: Initialize
[14]
Display at startup
Selects stand-by state or Counter display (origin
detection wait state for EG-101Z) to display at startup.
EG-101P/EG-101D
0: [------] display
1: 0.000
EG-101Z
0: [------] display
1:
Origin detection wait state
[16]
Calculation with a
constant
Sets whether to multiply the counter value by a
predetermined value, by an arbitrary value, or to
not multiply it.
The value obtained by multiplying the counter
value by the set constant value will be displayed as
the measurement result. For details about optional
constant value setting, see
Constant Value Setting" (page 6).
0: Do not calculate
2: 10 times
1: 2 times
3: Arbitrary value
[17]
Hide the lowest-order
digit*
2
Hides the lowest-order digit.
However, the lowest-order digit will be included in
printouts.
0: Display all digits
1:
Hide the lowest-order digit
[18]
Smoothing
(EG-101P/EG-101Z
only)
Averages the counter value and then displays it.
(This reduces fluctuation of the lowest-order digit.)
You can specify the number of measurements to
average. For EG-101D, this is not available.
0: None
1: Display the average of 8 measurements
2: Display the average of 16 measurements
[20]
Tolerance judgment/
BCD output
switchover
Switches between tolerance judgment result output
and BCD output.
0: Tolerance judgment result output
1: BCD output
[21]
Tolerance mode*
1
Selects the tolerance mode.
0: 3-step tolerance
1: 5-step tolerance
[22]
BCD output mode
Sets the output timing of BCD.
0: Command mode (sync control)
1: Interval mode
[23]
BCD output speed
Selects the speed of BCD.
0: 5 ms
2: 20 ms
1: 15 ms
3: 40 ms
[24]
BCD output logic
Selects the output logic of BCD.
0: DATA [H] (sign H)
2: DATA [H] (sign L)
1: DATA [L] (sign L)
3: DATA [L] (sign H)
[29]
Digimatic input
WAIT*
3
*
4
(EG-101D only)
Sets the wait time for the Digimatic input signal.
Change this when the Counter cannot read the
input signals from a Digimatic device.
0: No wait
2: 200 ms WAIT
1: 100 ms WAIT
[35]
Key protect
Key operations can be disabled to prevent
operation errors.
0: Key operation enabled
1: Key operation disabled
[41]
Origin detection
direction
(EG-101Z only)
When a Linear Gage with an origin mark is
connected, selects the direction of the spindle of
the Linear Gage for origin detection. For EG-101P,
this is not available.
0: + direction
1: - direction
[42]
Origin re-detection*
5
(EG-101Z only)
When a Linear Gage with an origin mark is
connected, sets whether to wait for the origin to
be detected without turning off the power in the
case of an abnormal stop. For EG-101P, this is not
available.
0: Disabled
1: Enabled
[Parameter number]
/Setting item
Description
(the values in bold indicate the default value)
[43]
Origin initialization
(when the power is
turned on)
(EG-101Z only)
When a Linear Gage with an origin mark is
connected, initializes the origin when the power
is on. After the initialization, the set value will be
returned to 0 (do not initialize). For EG-101P, this is
not available.
0: Do not initialize
1: Initialize
*
1
The Preset value and tolerance value that had been set will be cleared if the
setting is changed.
*
2
When this parameter is set to 1 "Hide the lowest-order digit", the "MIN"
display of the Peak mode indicator also becomes hidden.
Set up the Peak mode while selecting 0 "Display all digits" for this parameter.
Even if it is set to "Hide the lowest-order digit" later, the mode setting for the
Peak mode continues to be valid.
*
3
The display speed can be changed.
*
4
When using EG-101D, an error may occur when a special gage is
connected. In that case, set the value to 1 or 2. When using EG-101P, set
the value to 0.
*
5
When the setting is enabled, the Counter will wait for the origin re-detection
either after the Preset value or tolerance value is set, or when the HOLD
signal is raised. If the HOLD signal is input again during origin re-detection,
the origin re-detection function will be canceled (except during error
detection).
4
Basic Operations
4.1
Preset
The current value of the Counter can be set to an arbitrary value at any point
within the Linear Gage measuring range.
Tips
• Switch the current Peak mode setting to the normal measurement in advance.
For details, see
"4.2 Peak Mode Setting" (page 5).
• The factory default setting of the Preset value is 0.
•
To set the current value of the Counter to 0 after you have modified the Preset
value, set the Preset value to 0. The maximum value, minimum value, and TIR
value that have been set in Peak mode will be set to 0 at this time.
•
For the EG-101D (gage type is set to "ABS") and EG-101Z (origin detection
function is enabled) Counters, the effective Preset count is one million times.
As an example, the procedure for presetting the datum to 10.005 mm is
explained.
1
Repeatedly press [BANK] to select BANK 0.
»
The BANK indicator turns off.
Tips
While [BANK] is pressed down, the currently selected BANK number
is displayed. When the key is released, the Counter will return to the
Counter display.
2
Press [Fn] to switch to setup mode.
» The previous Preset value will be
displayed. (The example on the right
shows the previous value as 10.000.)
3
Press [MODE].
» The input digit will shift to the right. (The
currently selected digit will blink.)
4
Press [P.SET].
»
The Preset value will be modified.
Tips
• The ±
sign is also set at the most significant digit. To set the Preset value to a
negative value, repeatedly press [P.SET] until the Sign indicator lights.
•
To cancel the input, press [Fn]. The Counter will return to the Counter display.
5
Repeat step
3
and step
4
until the least significant digit has
been set.
»
The least significant digit will blink.
6
Press [MODE].
» The Preset value will be applied (the
least significant digit stops blinking).