Mitutoyo EG-101P Скачать руководство пользователя страница 4

4

Printed in Japan

Date of publication: February 1, 2019

[Parameter number] 

/Setting item

Description 

(the values in bold indicate the default value)

[10]

 

Parameter 

initialization*

1

If you set the value of this parameter to 1, the 

set values for all parameters, except for the 

unit setting, can be reset to their default values 

(initialized).  

Once this setting has been enabled, this parameter 

is reset so its set value is 0 (do not initialize).

0: Do not initialize

1: Initialize

[14]

 

Display at startup

Selects stand-by state or Counter display (origin 

detection wait state for EG-101Z) to display at startup.

EG-101P/EG-101D

0: [------] display

1: 0.000

EG-101Z

0: [------] display

1: 

Origin detection wait state

[16]

 

Calculation with a 

constant

Sets whether to multiply the counter value by a 

predetermined value, by an arbitrary value, or to 

not multiply it.  

The value obtained by multiplying the counter 

value by the set constant value will be displayed as 

the measurement result. For details about optional 

constant value setting, see 

"4.5 Optional 

Constant Value Setting" (page 6).

0: Do not calculate

2: 10 times

1: 2 times

3: Arbitrary value

[17]

 

Hide the lowest-order 

digit*

2

Hides the lowest-order digit.  

However, the lowest-order digit will be included in 

printouts.

0: Display all digits

1: 

Hide the lowest-order digit

[18]

 

Smoothing 

(EG-101P/EG-101Z 

only)

Averages the counter value and then displays it. 

(This reduces fluctuation of the lowest-order digit.) 

You can specify the number of measurements to 

average. For EG-101D, this is not available.

0: None

1: Display the average of 8 measurements
2: Display the average of 16 measurements

[20]

 

Tolerance judgment/

BCD output 

switchover

Switches between tolerance judgment result output 

and BCD output.

0: Tolerance judgment result output

1: BCD output

[21]

 

Tolerance mode*

1

Selects the tolerance mode.

0: 3-step tolerance

1: 5-step tolerance

[22]

 

BCD output mode

Sets the output timing of BCD.

0: Command mode (sync control)

1: Interval mode

[23]

 

BCD output speed

Selects the speed of BCD.

0: 5 ms

2: 20 ms

1: 15 ms

3: 40 ms

[24]

 

BCD output logic

Selects the output logic of BCD.

0: DATA [H] (sign H)

2: DATA [H] (sign L)

1: DATA [L] (sign L)
3: DATA [L] (sign H)

[29]

 

Digimatic input 

WAIT*

3

*

4

 

(EG-101D only)

Sets the wait time for the Digimatic input signal. 

Change this when the Counter cannot read the 

input signals from a Digimatic device.

0: No wait

2: 200 ms WAIT

1: 100 ms WAIT

[35]

 

Key protect

Key operations can be disabled to prevent 

operation errors.

0: Key operation enabled

1: Key operation disabled

[41]

 

Origin detection 

direction 

(EG-101Z only)

When a Linear Gage with an origin mark is 

connected, selects the direction of the spindle of 

the Linear Gage for origin detection. For EG-101P, 

this is not available.

0: + direction

1: - direction

[42]

 

Origin re-detection*

5

 

(EG-101Z only)

When a Linear Gage with an origin mark is 

connected, sets whether to wait for the origin to 

be detected without turning off the power in the 

case of an abnormal stop. For EG-101P, this is not 

available.

0: Disabled

1: Enabled

[Parameter number] 

/Setting item

Description 

(the values in bold indicate the default value)

[43]

 

Origin initialization 

(when the power is 

turned on) 

(EG-101Z only)

When a Linear Gage with an origin mark is 

connected, initializes the origin when the power 

is on. After the initialization, the set value will be 

returned to 0 (do not initialize). For EG-101P, this is 

not available.

0: Do not initialize

1: Initialize

*

1

     The Preset value and tolerance value that had been set will be cleared if the 

setting is changed.

*

2

   When this parameter is set to 1 "Hide the lowest-order digit", the "MIN" 

display of the Peak mode indicator also becomes hidden. 

Set up the Peak mode while selecting 0 "Display all digits" for this parameter. 

Even if it is set to "Hide the lowest-order digit" later, the mode setting for the 

Peak mode continues to be valid.

*

3

   The display speed can be changed.

*

4

   When using EG-101D, an error may occur when a special gage is 

connected. In that case, set the value to 1 or 2. When using EG-101P, set 

the value to 0.

*

5

   When the setting is enabled, the Counter will wait for the origin re-detection 

either after the Preset value or tolerance value is set, or when the HOLD 

signal is raised. If the HOLD signal is input again during origin re-detection, 

the origin re-detection function will be canceled (except during error 

detection).

Basic Operations

4.1 

Preset

The current value of the Counter can be set to an arbitrary value at any point 

within the Linear Gage measuring range.

Tips

•  Switch the current Peak mode setting to the normal measurement in advance. 

For details, see 

"4.2 Peak Mode Setting" (page 5).

•  The factory default setting of the Preset value is 0.
• 

To set the current value of the Counter to 0 after you have modified the Preset 

value, set the Preset value to 0. The maximum value, minimum value, and TIR 

value that have been set in Peak mode will be set to 0 at this time.

• 

For the EG-101D (gage type is set to "ABS") and EG-101Z (origin detection 

function is enabled) Counters, the effective Preset count is one million times.

As an example, the procedure for presetting the datum to 10.005 mm is 

explained.

Repeatedly press [BANK] to select BANK 0.

 

»

The BANK indicator turns off.

Tips

While [BANK] is pressed down, the currently selected BANK number 

is displayed. When the key is released, the Counter will return to the 

Counter display.

Press [Fn] to switch to setup mode.

 

» The previous Preset value will be 

displayed. (The example on the right 

shows the previous value as 10.000.)

Press [MODE].

 

» The input digit will shift to the right. (The 

currently selected digit will blink.)

Press [P.SET].

 

»

The Preset value will be modified.

Tips

•  The ±

 sign is also set at the most significant digit. To set the Preset value to a 

negative value, repeatedly press [P.SET] until the Sign indicator lights.

• 

To cancel the input, press [Fn]. The Counter will return to the Counter display.

Repeat step 

3

 and step 

4

 until the least significant digit has 

been set.

 

»

The least significant digit will blink.

Press [MODE].

 

» The Preset value will be applied (the 

least significant digit stops blinking).

Содержание EG-101P

Страница 1: ...nted Precautions for Use Do not remove the cover or disassemble the product Otherwise you may be subject to electric shock and there is a risk of breakage or fire due to a short circuit caused by meta...

Страница 2: ...ear Gage OUTPUT connector I O connector For connecting an I O connecting cable Grounding terminal For connecting a grounding wire Power terminal block For connecting the Terminal strip connecting cabl...

Страница 3: ...number to 12 The current value of parameter number 12 will blink Parameter number 12 sets the resolution Parameter number Set value 5 Repeatedly press P SET to set the set value to 1 resolution 5 m T...

Страница 4: ...p For EG 101P this is not available 0 Disabled 1 Enabled Parameter number Setting item Description the values in bold indicate the default value 43 Origin initialization when the power is turned on EG...

Страница 5: ...tolerance judgment function is disabled when BANK0 is selected 4 4 Tolerance Value Setting There are 2 settings for the tolerance value 3 step and 5 step Tips Set the value of parameter number 21 to...

Страница 6: ...st significant digit stops blinking 8 Press and hold Fn and then press P SET The Counter will return to the Counter display 5 External Input Output Function This product has an I O connector that enab...