Tips

For details about I/O connecting cables, see 

"5 External Input/Output 

Function" (page 6).

 

„

Connection procedure

• 

When making connections, connect the power cable last.

•  Do not run the power cable and Linear Gage connecting cable through a cable 

duct together with other power lines.

•  Secure the power cable and connecting cables for external equipment to your 

equipment with a cable tie, cable holder, etc.

Be sure to connect this product to ground. If this product is not grounded, 

it will be more susceptible to electrical noise.

Make connections as shown in the figure below.

Linear Gage 

input connector

Linear Gage

Grounding terminal

Grounding wire 

(Sold separately)

AC adapter 

(option)

Power

Terminal 

block

Terminal strip connecting cable 

(option)

2.4 

Operation Check

Check the cable connections with the following procedure to confirm that the 

connections are correct.

Connect the power.

 

» The Counter enters the stand-by state.

Press [P.SET].

 

» The Counter changes to the Counter 

display.

Tips

EG-101Z will enter the origin-detection waiting state. To change to the Counter 

display, push in the contact point of the Linear Gage to make it pass over the origin.

Check that the counter value is shown on the Display.

Check that the counter value on the Counter changes by moving 
the contact point of the Linear Gage up and down.

Parameter setting

The settings of the Linear Gage that you will use, the display of the 

Counter, and external output are specified by setting parameters. Set 

parameters before you begin measuring.

3.1 

Procedure for Setting Parameters

Parameters are set in Parameter mode. As an example, the procedure for 

using the Linear Gage with a resolution of 5 µm for EG-101P is explained.

Connect the power.

 

» The Counter enters the stand-by state.

Press and hold [Fn], and then press [P.SET].

 

» The Counter enters Parameter mode. 

(The set value of parameter number 00 

will blink.)

Parameter number

Set value

Press [P.SET] once to set the value to 1 (parameter setting).

 

»

Parameters can now be modified. (The 

set value remains blinking.)

Parameter number

Set value

Tips

If the setting value is 0, you can view the parameter values, but you cannot change 

them.

Repeatedly press [Fn] to advance the parameter number to 12.

 

» The current value of parameter number 

12 will blink. (Parameter number 12 

sets the resolution.)

Parameter number

Set value

Repeatedly press [P.SET] to set the set value to 1 (resolution: 
5 µm).

 

»

The value will be set to 1. (The Linear 

Gage resolution will be set to 5 µm.)

Parameter number

Set value

Press and hold [Fn], and then press [P.SET].

 

» The Counter will return to the stand-by 

state.

3.2 

Basic Parameters

This section explains the basic parameters related to measuring. Be sure to 

set these settings before measuring.

Tips

Correct measurement results may not be obtained if the settings are incorrect.

[Parameter number] 

/Setting item

Description 

(the values in bold indicate the default value)

[00]

 

Parameter mode

Used to view or modify parameters.

0: View parameters

1: Set parameters
2: Set an optional constant value*

1

[05]

 

Origin detection 

function*

2

 

(EG-101Z only)

Selects whether the origin is restored when a 

Linear Gage with an origin point mark is connected. 

For EG-101P, this is not available.

0: Disabled

1: Enabled

[11]

 

Counter direction

Sets whether the numeral will increase or decrease 

when the spindle of the Linear Gage is pushed in.

0: + direction

1: - direction

[12]

 

Linear Gage 

resolution

*

3

Sets the resolution or the type of the Linear Gage 

to be connected.

EG-101P/EG-101Z
0: 10 µm
1: 5 µm

2: 1 µm

3: 0.5 µm
4: 0.1 µm
5: 0.1 µm

EG-101D*

4

0: INC

1: ABS

[15]

 

Unit system 

selection*

3

The unit for displayed values can be set to "mm" 

or "E units". E=1/25.4 mm. After the unit is set, 

the default value will not be restored even if the 

parameters are re-initialized.

0: mm

1: E 5/100,000 reading
2: E 1/10,000 reading

3:    mm

 (when connecting an E gage, 1/10,000 reading. 

EG-101D only)

*

1

   The optional constant value setting is available only when the value of 

parameter number 16 is set to 3. For details, see 

"4.5 Optional Constant 

Value Setting" (page 6).

*

2

   To use the origin detection function with EG-101Z, set the value to 1. When 

using EG-101P, set the value to 0.

*

3

   The Preset value and tolerance value that had been set will be cleared if the 

setting is changed.

*

4

   An ABS type gage stores the origin even when the power is off. Set this 

according to the Linear Gage type. To match the display of the Counter and 

Linear Gages such as ID or SD, specify INC mode.

3.3 

Advanced Parameters

This section explains the parameters related to the display, functions, and 

external output of the Counter. Configure the settings appropriate to your 

application.

(exclusive for 542-711-1 and 542-712-1)

Содержание EG-101P

Страница 1: ...nted Precautions for Use Do not remove the cover or disassemble the product Otherwise you may be subject to electric shock and there is a risk of breakage or fire due to a short circuit caused by meta...

Страница 2: ...ear Gage OUTPUT connector I O connector For connecting an I O connecting cable Grounding terminal For connecting a grounding wire Power terminal block For connecting the Terminal strip connecting cabl...

Страница 3: ...number to 12 The current value of parameter number 12 will blink Parameter number 12 sets the resolution Parameter number Set value 5 Repeatedly press P SET to set the set value to 1 resolution 5 m T...

Страница 4: ...p For EG 101P this is not available 0 Disabled 1 Enabled Parameter number Setting item Description the values in bold indicate the default value 43 Origin initialization when the power is turned on EG...

Страница 5: ...tolerance judgment function is disabled when BANK0 is selected 4 4 Tolerance Value Setting There are 2 settings for the tolerance value 3 step and 5 step Tips Set the value of parameter number 21 to...