![Mitutoyo EG-101P Скачать руководство пользователя страница 8](http://html1.mh-extra.com/html/mitutoyo/eg-101p/eg-101p_user-manual_1812410008.webp)
8
Printed in Japan
Date of publication: February 1, 2019
Cause
Solution
Symptom 3: The Counter executes Zero setting unexpectedly.
A power interruption (sudden
drop in power voltage) has
occurred.
Check the power supply. For details
about how to cancel the error, see
[------] in the List of Error Codes.
"6.2 List of Error Codes" (page 8)
Symptom 4: BCD output is not working.
BCD output is not enabled.
Enable BCD output. (Set the value of
parameter number 20 to 1.)
Output logic is not correct.
Select the output logic that is
appropriate for your application by
setting parameter number 24.
Command mode is active.
Enable interval mode. (Set the value
of parameter number 22 to 1.)
Symptom 5: Cannot modify the parameter settings.
"Set parameters" (set value 1)
has not been set for Parameter
mode.
Set the value for parameter number
00 to 1. For details, see
Procedure for Setting Parameters"
6.2
List of Error Codes
Display NOM
signal
BCD
output
Cause
Solution/
Error cancellation
method
Err 10
H
FFFF10
Abnormal power
voltage
Connect to the specified
power.
Automatic cancellation
[------]
H
FFFF15
In stand-by state
after power-
on or a power
interruption
Re-check the power if a
power interruption has
occurred.
y
Press [P.SET].
y
Input an external HOLD
signal.
Err 20
H
FFFF20
Excess speed
Revise the measurement
conditions.
y
Press [P.SET].
y
Input an external HOLD
signal.
Err 30
H
FFFF30 Counter value is
8 digits or more
Modify the Preset value.
y
Press [P.SET].
y
Input an external HOLD
signal.
Err 40
H
FFFF40 Linear Gage
malfunction or
excess speed
y
Check the Linear Gage
connection.
y
Revise the
measurement
conditions.
y
Press [P.SET].
y
Input an external HOLD
signal.
F*****
L
F*****
Counter value is
6 digits or more
Return the counter value to
within 6 digits.
All
decimal
points
blinking
L
Counter
value
status
The origin is not
detected yet
(EG-101Z only)
Pass the spindle of the
Linear Gage through the
origin.
Automatic cancellation
Err 90
L
Counter
value
status
Tolerance value
setting error
Input the tolerance value
again.
Press [P.SET].
Err 95
L
Counter
value
status
Key protect
Cancel the key protection
parameter. (Set parameter
number 35 to 0.)
Automatic cancellation
Tips
•
If the NOM signal is "H": When an error occurs in 3-step or 5-step tolerance
judgment output (the value for parameter number 20 is 0), both pin numbers 3
(L1) and 7 (L5) will be "L".
•
If an error occurs while you are setting parameters, the Preset value, or the
tolerance value, the error will be displayed after you return to the counter state.
However, the error signal will be output immediately to any external output.
• Errors can also be canceled with an external PSET signal.
•
For EG-101D, all decimal points blink for about 8 seconds when an error is
canceled.
7
Specifications
Basic specifications
Code No.
542-015
542-017
542-016
Model No.
EG-101P
EG-101Z
EG-101D
Number of display axis
1 axis
Display
Minus (–) sign and 6 numeric digits (green LED)
Minimum reading
(Count display range)
Selected by parameter
Set
automatically
according to
the gage
0.01 (
±9999.99) mm
0.005 (
±
999.995) mm
0.001 (
±999.999) mm
0.0005 (
±
99.9995) mm
0.0001 (
±99.9999) mm
Maximum input
frequency
1.25 MHz
(2-phase square wave)
Maximum count speed
5 MHz
Power source voltage
DC +12 V to 24 V
Power consumption
Max. 6 W (Max. 500 mA)
Operating temperature
(humidity) range
0 °C to 40 °C
(20
%
RH to 80 % RH, without
condensation)
Storage temperature
(humidity) range
-10
°C
to 50
°C
(20 % RH to 80 % RH, without
condensation)
External dimensions
96 (W) ×
48 (H)
×
155 (D) mm
Mass
Approx. 400 g
CE marking
EMC directive EN61326-1
Immunity test requirement: Clause 6.2 Table 2
Emission limit: Class B
RoHS directive EN50581
Interface
I/O
External dimensions drawing (for all models)
154.6
48
96
44
.4
91.4
Unit: mm
Options
Part No.
Name
02ADB440
I/O output connector (with cover)
02ADD930
Terminal strip connecting cable*
1
357651
AC adapter
02ZAA000
AC cable*
1
*
1
Required if using the AC adapter.
Mitutoyo Corporation
20-1, Sakado, 1-Chome, Takatsu-ku, Kawasaki-shi,
Kanagawa 213-8533, Japan