41
E10
Communication Error
Turn the meter off, check the cables to
the computer, then turn the meter back
on. If error persists, contact your local
distributor.
E11
Ambient Temperature
out of range
Ambient temperature is too high or too
low to maintain a controlled test strip
temperature. Re-locate the meter into
the proper temperature environment.
Warning and Error Icons
Flashing ¼ full icon.
Batteries are
discharged and need
to be replaced soon
Test results will still be accurate, but
replace the batteries as soon as
possible.
Strip removed during
test
Insert a new test strip. Make sure to
wait until prompted to remove the
strip.
Low
battery
Low battery
Replace the batteries before the next
test is performed.
INR reading is >7.0
Repeat the test to ensure consistency
of readings. Run Electronic Calibrator
checks. Run liquid calibrator checks.
INR reading is <0.7
Repeat the test to ensure consistency
of readings. Run Electronic Calibrator
checks. Run liquid calibrator checks.
Code Chip problem
Missing or faulty code chip. Insert the
code chip found within the box of test
strips and/or remove and replace the
code chip.
Lid Left Open
Close the optics cover and restart the
test.