For each of the above four parameters it is possible to check the corresponding value
of the test with a personal value set by the doctor or a predicted value taken from a
predicted source, by selecting one of the two following items:
Personal best value setup
Use predicted source
Questions Setup
Specific questions can be written so that when the patient turns-on the device a table
will be displayed with a series of questions. The following table displays the items
which can be selected and the response options which can be used by the patient.
A multiple selection can also be implemented.
Symptoms Setup
Every time that a patient records a test the symptoms setup will have the patient
answer a series of questions; the following table displays the items which can be
selected and the response options that can be chosen by the patient:
Domanda
Possibili risposte
Tiredness on waking
NO
MED
MAX
Daytime drowsiness
NO
MED
MAX
Breathless on waking
NO
MED
MAX
Troubled sleep
NO
MED
MAX
Wheezing
NO
MED
MAX
Cough
NO
MED
MAX
Sputum production
NO
LIGKT
DARK
Sputum increasing
NO
YES
breathlessness
NO
Effort
At rest
Fatigue
NO
MED
MAX
Chest tightness
NO
MED
MAX
Also in this case a multiple selection can be implemented.
3.6.2
Reusable turbine calibration
WARNING
The turbine flow sensor does not require calibration, however regular
cleaning of the turbine is necessary . If a calibration must be performed the
following guidelines should be carefully noted.
A calibration can only be performed with the reusable turbine.
Calibration of the turbine is performed using a calibration syringe to make a FVC test
for the expired parameters and a FIVC test for the inspired parameters.
To access the calibration function, select the “Turbine Calibration” option from the
Service Menu as explained in paragraph 3.6. Once the password has been inserted
correctly the following screen will appear: