Appendix E. Dust filters.
The dust filters are available on special request. For the “Standard” and “Plastic” housings is
used
Top filter ESAT.717100.001
only (see Fig. 11
). For the “Fast response” modification is used
additional
Side filter ESAT.717100.002
(see Fig. 12). The filters material - Membrane
Владипор
МФФК ТУ 6-55-221-1413-2004
. On the dark areas applied
3M Double Linered Laminating
Adhesive 7952
.
Fig. 11. Top filter for the "Standard" and "Plastic" housings.
Fig. 12. Side filter for the "Fast response" housing.
To attach the dust filter, perform the following steps:
Wipe the sensor top surface with a cloth that is dampened with absolute alcohol.
For the
“Fast response” modification, perform the same operation on the side surface.
Retrieve the teflon tape-bonded dust filters from packing.
Detach a filter from the teflon tape with tweezers.
Align the filter to the top surface center with tweezers, and then press on it.
For the
“Fast response” modification, perform the same operation on the side surface.
After the dust filter was attached, zeroing procedure must be performed (see Appendix D for
details).
Perform this work at well illuminated and well ventilated place.
After attaching the dust filter, perform zeroing procedure (see Appendix
D for details).