![Mipex MIPEX-02 1 Series Скачать руководство пользователя страница 29](http://html1.mh-extra.com/html/mipex/mipex-02-1-series/mipex-02-1-series_user-manual_1804820029.webp)
OPTOSENSE LLC
Design Department
USER
MANUAL
SMALL-SIZE EXPLOSIVE GAS MEASURING SENSOR MIPEX-02-
Х-X-X.1 X
File name: ESAT.413347.006 UM v.2.0.docx
Revision 2.0 27 August 2015
Page. 29 of 38
THE SOLE PROPERTY OF OPTOSENSE LLC.
ANY REPRODUCTION
WITHOUT THE WRITTEN PERMISSION OF OPTOSENSE LLC IS PROHIBITED.
RT?
Yes
Check the type code (TYPE)
ASCII 5-byte
and 0Dh
The response includes the sensor type data (MIPEX model, target gas,
housing code and pinout code). The response format available by
request.
RX?
Yes
Check the characteristics code
(RX).
ASCII 2-byte
and 0Dh
RX codes according to Table 3.
SRAL?
Yes
Check the sensor serial number
(SN).
ASCII 8 byte
and 0Dh
SREV?
Yes
Check the firmware version.
Response includes the sensor model and current firmware version. For
example:
MIPEX-2_25.4
SREVH
No
History request of firmware
updates.
The command response is the 10 latest firmware versions. Example of
response
– see chapter C.4.
USERDATA?
No
Display all data stored in memory.
Stacked 10
digits shows
There is ability to save up to 10 digits (0-99999) to flash-memory of the
sensor. After firmware update from the 25.2 version, where this function
is absent, all default values are «
67294
».
USERDATA
XX
?
No
Check data in the sensor memory.
Value from
00 to 09
Display the data saved in cell with
ХХ
number.
USERDATA
XX
YYYYY
No
Save data to the sensor memory.
YYYYY
Save
YYYYY
value to cell number
ХХ.
NETON
No
Enable to storing network address
NETON OK
After turn off power, net address will be reserved.
NETOFF
No
Disable address storing.
NETOFF OK
After turn off power, net address will be erased (default 00 value).
!**
No
Check current network address.
XX
Response is network hexadecimal address
%XXYY
Yes
Set up network address.
YY
XX
– old address,
YY
– new address.
UART?
Yes
Request current work mode.
USER
or
OEM
UPLOAD
No
Switch the sensor to firmware
update bootloader mode.
U_OK
See chapter C.4 for details.