Table 4. Response time of the sensors.
Construction
Target gas
Max. response
time t(90), sec
stainless steel, “Standard” and
plastic
CH4
30
CH4/СН4+С2H6
C3H8
CO2
60
stainless steel, with side holes,
“Fast response”
CH4
10
CH4/СН4+С2H6
C3H8
CO2
20
Table 5. General sensor accuracy* depends on calibration gas
Calibration
gas
Accuracy over temperature
range
Additional pressure
error
Additional humidity
error
CH4
±0,1%Vol. or ±5% of
indication in range
+20…+25°C;
±0,2% Vol. or ±10% of
indication in range
–10…+20°C and
+25…+40°C;
±0,4% Vol. or ±20% of
indication in range
-
40…-10°C and +40…+60°C.
±0,2% Vol. or ±30% of
indication from
100kPa (test: 80kPa,
100kPa, 120kPa)
±0,2 % Vol. or ±15 %
of indication from the
indication at
adjustment at 40 °C
(test: 20 %RH, 50
%RH,90 %RH)
CO2, C3H8
±0,05% Vol. or ±5% of
indication in range
+20…+25°C;
±0,1% Vol. or ±10% of
indication in range
–10…+20°C and
+25…+40°C;
±0,2% Vol. or ±20% of
indication in range
-40
…-10°C and +40…+60°C.
±0,1% Vol. or ±30% of
indication from
100kPa (test: 80kPa,
100kPa, 120kPa)
±0,1 % Vol. or ±15 %
of indication from the
indication at
adjustment at 40 °C
(test: 20 %RH, 50
%RH,90 %RH)
*Table shows the general accuracy, but individual sensor accuracy depends on RX characteristic
and limited by the
temperature range
(see Table 3).