Table 9. Status word description
Status word
Status
byte L
(DATAE or
DATAE2)
Status
byte H
(DATAE or
DATAE2)
Description
Recommendation
00
Normal mode, static temperature
10
Bit 0
Warm-up
Do not perform calibration.
11
Bit 0
Too many UART interruptions
Do not perform calibration or zeroing. Decrease request frequency.
21
Bit 4
Fast temperature change
Do not perform calibration or zeroing.
22
Bit 5
Sharp temperature change
Do not perform calibration or zeroing.
24
Bit 4
Bit 1
Fast temperature change + Zero
shifts to negative values
Do not perform calibration or zeroing.
30
Bit 2
Low signals level
Do not perform calibration or zeroing.
Measuring accuracy need to be additionally verified, once
switching on of Bit 2 has been observed.
Replace the sensor if Bit2 is on constantly (> 20 min).
31
Bit 1
Zero shifts to negative values
Using of dust filters is recommended to avoid moisture or dust
inside sensor. In case of moisture ingress, unmounts sensor off
device, then dry it during 4 hours minimum in warm dry location.
Make zeroing procedure and performance check.
40
Bit 6
Out of operation temperature range
Check the ambient temperature.
50
Bit 1
Abrupt change of signal
Do not perform calibration or zeroing.
Replace sensor, if bit1 is on constantly (> 20 min).
51
Complex status
Contact support for details
90
Bit 7
Firmware corruption
Contact support for details
1XX
Bit 2
Low power mode is on. XX- status
word values similar for normal mode.
To achieve full compliance to declared specification turn on the
standard mode