Table 12. CGM used for CH4 calibration gas.
CGM Nos
according
to text
Component
composition
Composition
of measured
component,
vol. %,
(% of LEL)
Permissible
deviation
limits, vol. %
Limits of
permissible
error of
qualification,
vol. %
Number as per
State Register
or Standard
designation
1
N
2
100
-
-
GOST
9392-74
2
СН
4
– N
2
2,2 (50)
±0,25
±0,04
3883-87
3
СН
4
– N
2
4,15 (94)
±0,25
±0,04
3883-87
Table 13. CGM used for C3H8 calibration gas.
CGM Nos
according
to text
Component
composition
Composition
of measured
component,
vol. %,
(% of LEL)
Permissible
deviation
limits, vol. %
Limits of
permissible
error of
qualification,
vol. %
Number as per
State Register
or Standard
designation
1
N
2
100
-
-
GOST
9392-74
2
С
3
Н
8
– N
2
0,85 (50)
±0,05
±0,015
5328-90
3
С
3
Н
8
– N
2
1,6 (94)
±0,1
±0,05
EM 06.01.648
Table 14. CGM used for CO2 calibration gas.
CGM Nos
according
to text
Component
composition
Composition
of measured
component,
vol. %,
(% of LEL)
Permissible
deviation
limits, vol. %
Limits of
permissible
error of
qualification,
vol. %
Number as per
State Register
or Standard
designation
1
N
2
100
-
-
GOST 9392-74
2
СО
2
– N
2
1
±0,05
±0,015
3
СО
2
– N
2
2.5
±0,1
±0,05