C.3.2. Request
commands
of factory settings and properties.
The commands shown in Table 10 are used for sensor setup, check the factory settings and firmware update.
There is able to work with up to 256 sensors in one UART line. Use prefix
#XX
before any command to send commands to any sensor in
one UART line.
XX
is a network hexadecimal address from 00 to FF.
Table 10. Setting commands
Command in
OEM mode
Available in
USER mode
Command name
Response
Description
LOWPW
x
No
Setup power mode.
LOWPW
x
OK
x
= 0
– normal mode of power consumption (
HI
),
х
= 1
– low power
consumption mode (
LW
).
LOWPW?
No
Check the current sensor mode.
HI
or
LW
OEM
XXXX
Yes
Change mode from
USER
to
OEM.
USER
or
OEM
XXXX
is digital password. If the password not correct, sensor stay in
current mode.
USER
Yes
Change mode from
OEM
to
USER.
USER
PASS
XXXX
YYYY
No
Password change.
PASS
XXXX
YYYY OK
XXXX
is old password,
YYYY
is new password. By default the
password is
‘’
0000
’’.
DATEZC
DD.MM.YY
No
Write calibration date in format
DD.MM.YY.
DATEZC
DD.MM.YY
OK
DD
– day (1-31),
MM
– month (1-12),
YY
– year (00 – 99).
DATEZC?
Yes
Check calibration date.
DD.MM.YY
CRC
Yes
Firmware integrity check, calculated
according to CRC16 algorithm.
CRC OK
and
XXXXX
YYYYY OK
After time-out during the calculations, not more than 1 sec).
XXXXX
–
calculated checksum for current firmware,
YYYYY
– checksum stored in
memory of the sensor. In case of malfunction,
XXXXX
and
YYYYY
values does not match together. Status word indicates 90 value if CRC
mismatches.
ID?
Yes
Check identifier.
TYPE SN RX
SREV
Returns summary info about type code, serial number, RX code and
firmware version separated with space symbol.
INDSIG ON
No
Turn on negative shift indication in
concentration readings.
ON
The command activates the possibility to show concentration shift to
negative values. In case of using this command, response on
concentration request is
-2
(status word 31) or
-3
(status word 24) in
ASCII. It is not needed to activate mode after power reset.
INDSIG OFF
No
Turn off negative shift indication in
concentration readings.
OFF