2.3. Open/Short calibration:
LCR01A/B function with Open/Short calibration to meet the measuring of
High/Low impedance measurement。
●Open calibration
Clean the measuring terminal,make it on open state,press【CAL】button
about 1 second,main parameter will display「CALO」and a “bi” sound will
be heard,Calibration time will depend on measuring frequency,the high
frequency,the lower calibration time needed;After calibration, a “bi” sound
will be heard again。
●Short calibration
Plug-in shorted bar first,press【CAL】about 2 seconds,till「CALS」shown
on main display,and a “bi” sound will be heard;After calibration, a “bi”
sound will be heard again。