37
Append
2
Open/Short compensation
For those precision impedance measuring instrument,the open and short
compensation need to be used to reduce the parasitic effect of the test
fixture。The parasitic effect of the test fixture can be treated like the simple
passive components in figure 3.7a。When the DUT is open,the instrument gets
the conductance Yp=Gp+jwCp(Figure 3.7b) 。When the DUT is short,the
instrument gets the impedance Zs=Rs+jwLs(Figure 3.7c)。After open and
short compensation,Yp and Zs are for calculating the real Zdut(Figure 3.7d)。
Redundant Zs
H
CUR
●
Parastic Yo
D
U
T
Co
●
Rs Ls
●
●
●
H
POT
L
POT
L
CUR
Parastic effect of test fixture
Zm
●
Go
Parastic of test fixture