29
3-Terminals(3T) uses coaxial cable to reduce the effect of the
parasitic capacitor,the shield of the coaxial cable should be connect
to Guard of the instrument to increase the measurement range up to
10 M
Ω
。
H
CUR
●
A
V
D
U
T
Co
●
Ro Lo
Ro Lo
D
U
T
●
●
●
●
H
POT
L
POT
L
CUR
●
Co no effect on test result
Figure 2.