Page 45 of 50
AN-49-003 Rev.: E (September 6, 2019) M175867 (R95153) File: AN49003.doc
This document and its contents are the property of Mini-Circuits
3.10
Tracking Calibration status:
In the main screen, above the serial number is a
CAL
button
.
Clicking on it will open the Calibration tracking
window.
Figure 3.10
Calibration status button
3.10.1
Not all firmware revisions support
the tracking of operating time and calibration status.
See Table below for required firmware revision.
Note:
Upgrading from a firmware version which does not support operating time tracking to one which
does, will result in the operating time being counted from when the firmware was upgraded, instead
of from last calibration.
Model
SSG-
15G-RC
Earliest
supported
firmware
A2
All Rev.
All Rev.
3.10.2
The calibration tracking window
shows the operating time from last calibration and allows
calibration reminders to be scheduled based on either operating time (recommended interval 4160 hours) or
date (recommended interval 2 years).
Figure