7.3 Function description 

Vacuum sterilizer divided into the following process: preheating process;the 
vacuum step-up process; sterilization process; exhaust drying process; the 
pressure balance process. 

7.3.1 preheating process:   

Select the program firstly, then close the door, press start button to enter the 
preheating process,electrify heated steam generator heater reaches 100 

℃, 

electrify heated the chamber wall heater to set temperature 50 

℃.

 

7.3.2 vacuum process:   

Chamber wall temperature to 50 

℃, when the steam generator temperature 

reaches 100 

℃, into the vacuum heating process, open the vacuum selection valve, 

exhaust valve (normally on) and start the vacuum pump, 300 seconds later Close 
the vacuum selection valve, exhaust valve (normally on) and the vacuum pump to 
stop pumping the vacuum.    Pot inlet valve intermittent pumping power, as steam 
generator to generate steam.    If the system is set up one pre-vacuum, the pressure 
rise to +50 kPa until the pot stops pumping, open the drain valve to drain the 
exhaust 0.0kPa to close drain valve, intermittent steam generator to pump water to 
generate steam, when the pot the temperature reaches the set value, the end of the 
vacuum and transferred to sterilized itinerary.    If the system is set three times 
pre-vacuum, then when the pressure inside the chamber rise to +50kPa ,end the 
first pulse.    Start the pump again, begin the second pre-vacuum process.   

7.3.3 Sterilization process:   

This trip is insulation packing process, when the pot temperature dropped to the 
set sterilization temperature of +0.5 

℃ , the pump starts pumping water to the 

steam generator, the pot temperature rise to the set sterilization temperature of 
+0.9 

℃ ,open the exhaust exhaust valve, allowing the pot temperature is kept 

between the set sterilization tempe0.3 

℃ to the set sterilization 

tempe0.9 

℃.

   

7.3.4 Exhaust drying process:   

When the sterilization time, reduced to zero, into the exhaust stroke dry, open the 
drain valve to drain the pressure reduced to +0.0then the vacuum pump to start. 
When the drying time only 3 minutes ,open the intake valve to vacuum drying.   

7.3.5 Pressure equilibrium process:   

Содержание SUN12-II

Страница 1: ...on Manual Please read the manual carefully operate it according to the manual strictly and avoid any hazard This manual applies to SUN8 SUN12 SUN16 SUN18 SUN23 Steam Sterilizer File No MTQM C1 7 5 1 0...

Страница 2: ...am Specification 2 5 Installation 3 6 Packing accessory 5 7 Display and keypad description 5 8 Emergency Condition 8 9 Maintenance and Service 9 10 Transportation and Storage 11 11 Attention 12 Append...

Страница 3: ...n Scope This machine can be applied in sterilizing for the tooth device operation device Sanitary Material and other heat resisting articles existing in the Dept of Stomatology Ophthalmology and lab T...

Страница 4: ...temperature of operate 0 40 C Type of sterilization cycles Class B Wire heat resistance level 70 C N W 23L 53kg N W 18L 48kg N W 16L 47kg N W 12L 46kg N W 8L 44kg 4 Diagram Specification Symbol Meanin...

Страница 5: ...onment temperature 5 40 Relative humidity 85 barometric pressure 70kPa 106kPa minimum brightness 215 15 lx 5 3 The Steam Sterilizer shall be installed on the horizontal working base Put machine on the...

Страница 6: ...ay contact you local supply authority for further information Hinweis Beim Anlaufen Starten dieser Ausr stungen kann ein kurzzeitiger Spannungseinbruch auftreten insbesondere bei schlechter Netzqualit...

Страница 7: ...for displaying kinds of status of the autoclave during its working time See more details in Appendix working status display form When alarm occurs to the autoclave the digital pipe of the display will...

Страница 8: ...ndicator for alarming the shortage of water in the distilled water tank and reminding for filling distilled water 11 Light indicator waste water tank full It s the light indicator for alarming the ful...

Страница 9: ...umping open the drain valve to drain the exhaust 0 0kPa to close drain valve intermittent steam generator to pump water to generate steam when the pot the temperature reaches the set value the end of...

Страница 10: ...e provide accessories to take devices Must use handheld to take things prevent burns 8 Emergency Condition During the working time the Steam Sterilizer will automatically alarming then release pressur...

Страница 11: ...to the right MAX position 7 E7 long du Fault of temperature senor on steam generator Inspecting the temperature sensor on steam generator 8 E8 long du Fail on pressure rising Inspect water flowing ro...

Страница 12: ...ommendation by the supplier 9 5 The replacement work for sealing ring Tool One flat screwdriver No sharp edge at the head 1 One hand catches the lip of the sealing ring the other hand insert the screw...

Страница 13: ...al components must cut off power supply first And only the manufacturer or professional person designated by the manufacturer can operate 10 Transportation and Storage 10 1 The preparatory work before...

Страница 14: ...e door of the Steam Sterilizer in case of scald when open the door 9 Shut off the power before discharging installing the sealing ring and the work shall be carried out after further sufficient coolin...

Страница 15: ...13 Appendix 1 Steam loop Work Description...

Страница 16: ...14 2 Circuit diagram of sterilizer...

Страница 17: ...playing Ed B D test when put the test material onto the tray inside the chamber firstly turn on the power by the green switch close the door keeping press the UNWRAPPED button around 10s When the boar...

Отзывы: