11
2)When the pressure in the sterilizer reaches 100 kpa
,
Pull rings on safety
valve, make the safety valve open,about 2S, if steam jet, the safety valv
e work is normal. Ifnot, the safety valve need for effective inspection or
replacement
Pull rings on safety valve will steam jet, It is best to use tools such
as screwdriver when pull. Do not use finger directly, Operator also need ke
ep away from it , prevent burns!
Replace the safety valve:
This operation is limited to the professionals
1) Loosen the set screws of the safety valve, take the safety valve down f
rom the panel,
2) replace with a qualified safety valve (must provide by manufacturers).
3) Test a sterilization process
The machine maintenance and replacement of electrical components must cut
off power supply first. And only the manufacturer or professional person
designated by the manufacturer can operate
10. Transportation and Storage
10.1 The preparatory work before
transportation and storage:
Shut off the power, draw out the plug and cool the Steam Sterilizer down.
10.2 Draining
Empty the water storing tank after the machine become colling; use the pipe
packing together with the machine, connecting to the valve showing in the above
picture 4 or 3, 4 is the water exiting mouth for used water collecting tank, 3 is the
water exiting mouth of the distilled source water storing tank. Revolving the
water exit button by anticlockwise direction so as to release the water.
10.3 The Steam Sterilizer shall be transferred and stored with the below
conditions:
Ambiance temperature: -
40ºC
-
+55ºC
Relative humidity:
<
85%
Atmospheric pressure: 500HPA-1060HPA
Note: Do not drag or drop during transportation.