10 

9.1 Clean the top water tank by disinfectant or alcohol once a week. 
9.2  Disinfect  and  clean  the  interior  surface  of  the  Steam  Sterilizer  chamber 
monthly. 
9.3 Replace the fuse.   

Fuse Type: 250V,15A 

  Check the parameters of new fuse is correct . 

1)

 

Before change the fuse, turn off the power, and must pull out the plug. 

2)  Revolve  the  fuse  base  by  counter-clockwise  by  screwdriver  to  bring  out  the 
melting fuse to be replaced; 
3) Replace the new fuse and reset the fuse base on the original position, then use 
the screwdriver to fasten the base by clockwise. 
4)Check the correctness of the data of the new fuse replaced.   
 
9.4 Clean the sealing ring periodically 
The operator shall clean the sealing ring periodically for avoiding the influence to 
the  seal caused  by  the  dust/dirt  left  due to  the  long  period  use.  Use  the  smooth 
cloth  with  distilled  water  to  wipe  the  surface  of  the  sealing  ring  or  sealing  cap 
gently.  The  user  shall  discharge  the  sealing  ring  for  further  cleaning  or 
replacement if the air-leaking problem can not be resolved after the above process 
(The user shall discharge the sealing ring periodically for cleaning and inspection 
as recommendation by the supplier).       
 
9.5 The replacement work for sealing ring 
Tool: One flat screwdriver (No sharp edge at the head) 
1)  One  hand  catches  the  lip  of  the  sealing  ring,  the  other  hand  insert  the 
screwdriver into the clearance between the sealing ring and the door for raise up 
the sealing ring. 
2)After raising up part of the ring, you can use hand to draw the whole ring out. 
Wash its groove after drawing out the ring and pay attention to see whet

her it’s 

spoiled or not in consideration of the necessity of replacement. 
3)Put back the ring to the original door groove after cleaning. Most Important: the 
inset  work  must  be  done  equably to  the groove  During  the  installation,  the  laid 
four  equal  points  of  the  ring  must  be inserted  to the  groove  firstly, later  for  the 
same  work  for  the  remaining  segments  of  the  ring.  After  finishing,  press  the 
sealing ring equally by the force of hand. 
4)Note: The inner circle of the ring may be raised up when insert the ring into the 
groove. Please press it to the groove by using the screwdriver carefully. 
Shut off the power and cooling the  Steam Sterilizer sufficiently before inserting 
the ring for avoiding scald. 
9.6 Check the safety valve 
In order to prevent the safety valve in the blocking state, Under normal use, allow 
the steam pressure release through it once a month.   
1)operation  sterilizing  procedure   

Содержание SUN12-II

Страница 1: ...on Manual Please read the manual carefully operate it according to the manual strictly and avoid any hazard This manual applies to SUN8 SUN12 SUN16 SUN18 SUN23 Steam Sterilizer File No MTQM C1 7 5 1 0...

Страница 2: ...am Specification 2 5 Installation 3 6 Packing accessory 5 7 Display and keypad description 5 8 Emergency Condition 8 9 Maintenance and Service 9 10 Transportation and Storage 11 11 Attention 12 Append...

Страница 3: ...n Scope This machine can be applied in sterilizing for the tooth device operation device Sanitary Material and other heat resisting articles existing in the Dept of Stomatology Ophthalmology and lab T...

Страница 4: ...temperature of operate 0 40 C Type of sterilization cycles Class B Wire heat resistance level 70 C N W 23L 53kg N W 18L 48kg N W 16L 47kg N W 12L 46kg N W 8L 44kg 4 Diagram Specification Symbol Meanin...

Страница 5: ...onment temperature 5 40 Relative humidity 85 barometric pressure 70kPa 106kPa minimum brightness 215 15 lx 5 3 The Steam Sterilizer shall be installed on the horizontal working base Put machine on the...

Страница 6: ...ay contact you local supply authority for further information Hinweis Beim Anlaufen Starten dieser Ausr stungen kann ein kurzzeitiger Spannungseinbruch auftreten insbesondere bei schlechter Netzqualit...

Страница 7: ...for displaying kinds of status of the autoclave during its working time See more details in Appendix working status display form When alarm occurs to the autoclave the digital pipe of the display will...

Страница 8: ...ndicator for alarming the shortage of water in the distilled water tank and reminding for filling distilled water 11 Light indicator waste water tank full It s the light indicator for alarming the ful...

Страница 9: ...umping open the drain valve to drain the exhaust 0 0kPa to close drain valve intermittent steam generator to pump water to generate steam when the pot the temperature reaches the set value the end of...

Страница 10: ...e provide accessories to take devices Must use handheld to take things prevent burns 8 Emergency Condition During the working time the Steam Sterilizer will automatically alarming then release pressur...

Страница 11: ...to the right MAX position 7 E7 long du Fault of temperature senor on steam generator Inspecting the temperature sensor on steam generator 8 E8 long du Fail on pressure rising Inspect water flowing ro...

Страница 12: ...ommendation by the supplier 9 5 The replacement work for sealing ring Tool One flat screwdriver No sharp edge at the head 1 One hand catches the lip of the sealing ring the other hand insert the screw...

Страница 13: ...al components must cut off power supply first And only the manufacturer or professional person designated by the manufacturer can operate 10 Transportation and Storage 10 1 The preparatory work before...

Страница 14: ...e door of the Steam Sterilizer in case of scald when open the door 9 Shut off the power before discharging installing the sealing ring and the work shall be carried out after further sufficient coolin...

Страница 15: ...13 Appendix 1 Steam loop Work Description...

Страница 16: ...14 2 Circuit diagram of sterilizer...

Страница 17: ...playing Ed B D test when put the test material onto the tray inside the chamber firstly turn on the power by the green switch close the door keeping press the UNWRAPPED button around 10s When the boar...

Отзывы: