  Steam Sterilizer 

1

General Introduction 

This Steam Sterilizer is a precision instrument specializing in the operation 
occasion in need of simple, fast and efficient sterilization by the doctor or his/her 
assistants, and the doctor shall take the responsibility for such work.   
This  Steam  Sterilizer  is  specially  designed  for  the  occasions  in  use  of  frequent 
sterilizations.  It  is  equipped  with  the  microprocessor  for  intelligent  control  and 
man-machine  interface  for  convenient  operation.  With  the  distinctions  in 
displaying parameter and condition actively, evaluating malfunction automatically, 
shutting off for self-protection automatically in case of excessive temperature and 
pressure, this type Steam Sterilizer is reliable product in sterilizing.

         

 
 
 
 
 
                                                                                                           
                                                                                                           
 

                     

 

2

Application Scope 

This machine can be applied in sterilizing for the tooth device, operation device, 

Sanitary Material

 and other heat-resisting articles existing in the Dept. of 

Stomatology, 

Ophthalmology and lab. 

This  product  is  not  suitable  for 

leakproofness.

  n

ot  resistant  to  high  temperature 

products,the sterilizer max 

temperature is 138 .0

This product is not suitable for  liquid sterilizing. 

 

 

3

Tech. Parameter 

Chamber size of the Steam Sterilizer(Diameter X Depth) 8L

 170mmX310mm 

 

Chamber size of the Steam Sterilizer (Diameter X Depth) 12L

200mmX360mm   

Chamber size of the Steam Sterilizer (Diameter X Depth) 16L

230mmX360mm 

Chamber size of the Steam Sterilizer (Diameter X Depth) 18L

249mmX355mm   

Chamber size of the Steam Sterilizer (Diameter X Depth) 23L

249mmX460mm 

Rated Voltage

………………………………………

230

, 50Hz

 

 

Rated Power

………………………………………

1800VA

           

2300VA 

1  Air Filter 

2  Safety valve  3  Air outlet  4  Power 

socket 

Fuse 

6  Distilled 

water outlet 

7  Used water  

outlet 

8  Filter of 

chamber  

9  Switch   10  Distilled 

Water inlet 

 

 

Содержание SUN12-II

Страница 1: ...on Manual Please read the manual carefully operate it according to the manual strictly and avoid any hazard This manual applies to SUN8 SUN12 SUN16 SUN18 SUN23 Steam Sterilizer File No MTQM C1 7 5 1 0...

Страница 2: ...am Specification 2 5 Installation 3 6 Packing accessory 5 7 Display and keypad description 5 8 Emergency Condition 8 9 Maintenance and Service 9 10 Transportation and Storage 11 11 Attention 12 Append...

Страница 3: ...n Scope This machine can be applied in sterilizing for the tooth device operation device Sanitary Material and other heat resisting articles existing in the Dept of Stomatology Ophthalmology and lab T...

Страница 4: ...temperature of operate 0 40 C Type of sterilization cycles Class B Wire heat resistance level 70 C N W 23L 53kg N W 18L 48kg N W 16L 47kg N W 12L 46kg N W 8L 44kg 4 Diagram Specification Symbol Meanin...

Страница 5: ...onment temperature 5 40 Relative humidity 85 barometric pressure 70kPa 106kPa minimum brightness 215 15 lx 5 3 The Steam Sterilizer shall be installed on the horizontal working base Put machine on the...

Страница 6: ...ay contact you local supply authority for further information Hinweis Beim Anlaufen Starten dieser Ausr stungen kann ein kurzzeitiger Spannungseinbruch auftreten insbesondere bei schlechter Netzqualit...

Страница 7: ...for displaying kinds of status of the autoclave during its working time See more details in Appendix working status display form When alarm occurs to the autoclave the digital pipe of the display will...

Страница 8: ...ndicator for alarming the shortage of water in the distilled water tank and reminding for filling distilled water 11 Light indicator waste water tank full It s the light indicator for alarming the ful...

Страница 9: ...umping open the drain valve to drain the exhaust 0 0kPa to close drain valve intermittent steam generator to pump water to generate steam when the pot the temperature reaches the set value the end of...

Страница 10: ...e provide accessories to take devices Must use handheld to take things prevent burns 8 Emergency Condition During the working time the Steam Sterilizer will automatically alarming then release pressur...

Страница 11: ...to the right MAX position 7 E7 long du Fault of temperature senor on steam generator Inspecting the temperature sensor on steam generator 8 E8 long du Fail on pressure rising Inspect water flowing ro...

Страница 12: ...ommendation by the supplier 9 5 The replacement work for sealing ring Tool One flat screwdriver No sharp edge at the head 1 One hand catches the lip of the sealing ring the other hand insert the screw...

Страница 13: ...al components must cut off power supply first And only the manufacturer or professional person designated by the manufacturer can operate 10 Transportation and Storage 10 1 The preparatory work before...

Страница 14: ...e door of the Steam Sterilizer in case of scald when open the door 9 Shut off the power before discharging installing the sealing ring and the work shall be carried out after further sufficient coolin...

Страница 15: ...13 Appendix 1 Steam loop Work Description...

Страница 16: ...14 2 Circuit diagram of sterilizer...

Страница 17: ...playing Ed B D test when put the test material onto the tray inside the chamber firstly turn on the power by the green switch close the door keeping press the UNWRAPPED button around 10s When the boar...

Отзывы: