MingTai SUN12-II Скачать руководство пользователя страница 11

 

 

Error code and resolving measures

 

NO  Code   

display 

Alarm Tone    Malfunction content 

Resolving Measures 

E1 

long "du" 
 

Fault of temperature   
sensor inside the chamber 

Inspect if anything effected the 
sensor or wire not connecting well 
or sensor was broken 

E2 

long "du" 

Pressure over 240 KPa 

Inspect pressure sensor or vacuum   
pump working 

E3 

long "du" 

Fault of temperature sensor outside 
the chamber 

Inspect the temperature sensor 
outside of the chamber 

E4 

long "du" 

Super high inner temperature when 
pressure rising 

Inspect inside temperature sensor 
or pressure sensor 

E5 

long "du" 

Pressure releasing slowly 

Inspect the filter inside the chamber 
keep it clean and no jam 

E6 

long "du" 

Problem of door switch or door 
being opened when working 

Inspect door switch or if door 
handle were revolved to the   
right MAX position 

E7 

long "du" 

Fault of temperature senor on steam 
generator 

Inspecting  the  temperature  sensor 
on steam generator 

E8 

long "du" 

Fail on pressure rising 

Inspect water flowing road and 
steam leaking. 


 

EH 

long "du" 

Steam generator not heating 

Inspect connecting wire or 
resistance of steam generator 

10 

EF 

long "du" 

Pressure not over than 0 KPa 

Inspect water pump/steam generator   

11 

EL 

long "du" 

Pressure not over than 20 KPa 

Inspect  if  electric  valve  causing 
steam leak, or replace water pump   

12 

Eo 

long "du" 

Pressure not over than 65 KPa 

Inspect if the water pump working 
weakly. Or replace it 

13 

E9 

long "du" 

Fail on keeping pressure and 
temperature 

Inspect the steam leaking 

14 

EC 

long "du" 

The process interrupted artificially 
 

Long press the start key, can unlock 
alarm. 
 

 

9. Maintenance and Service 

 

Shut  off  the  power  before  service  or  components’  replacement,  and  the 

service or replacement work shall be done by the technician. 
 
Working environment minimum brightness to 215 + 15 lx. 
 

Содержание SUN12-II

Страница 1: ...on Manual Please read the manual carefully operate it according to the manual strictly and avoid any hazard This manual applies to SUN8 SUN12 SUN16 SUN18 SUN23 Steam Sterilizer File No MTQM C1 7 5 1 0...

Страница 2: ...am Specification 2 5 Installation 3 6 Packing accessory 5 7 Display and keypad description 5 8 Emergency Condition 8 9 Maintenance and Service 9 10 Transportation and Storage 11 11 Attention 12 Append...

Страница 3: ...n Scope This machine can be applied in sterilizing for the tooth device operation device Sanitary Material and other heat resisting articles existing in the Dept of Stomatology Ophthalmology and lab T...

Страница 4: ...temperature of operate 0 40 C Type of sterilization cycles Class B Wire heat resistance level 70 C N W 23L 53kg N W 18L 48kg N W 16L 47kg N W 12L 46kg N W 8L 44kg 4 Diagram Specification Symbol Meanin...

Страница 5: ...onment temperature 5 40 Relative humidity 85 barometric pressure 70kPa 106kPa minimum brightness 215 15 lx 5 3 The Steam Sterilizer shall be installed on the horizontal working base Put machine on the...

Страница 6: ...ay contact you local supply authority for further information Hinweis Beim Anlaufen Starten dieser Ausr stungen kann ein kurzzeitiger Spannungseinbruch auftreten insbesondere bei schlechter Netzqualit...

Страница 7: ...for displaying kinds of status of the autoclave during its working time See more details in Appendix working status display form When alarm occurs to the autoclave the digital pipe of the display will...

Страница 8: ...ndicator for alarming the shortage of water in the distilled water tank and reminding for filling distilled water 11 Light indicator waste water tank full It s the light indicator for alarming the ful...

Страница 9: ...umping open the drain valve to drain the exhaust 0 0kPa to close drain valve intermittent steam generator to pump water to generate steam when the pot the temperature reaches the set value the end of...

Страница 10: ...e provide accessories to take devices Must use handheld to take things prevent burns 8 Emergency Condition During the working time the Steam Sterilizer will automatically alarming then release pressur...

Страница 11: ...to the right MAX position 7 E7 long du Fault of temperature senor on steam generator Inspecting the temperature sensor on steam generator 8 E8 long du Fail on pressure rising Inspect water flowing ro...

Страница 12: ...ommendation by the supplier 9 5 The replacement work for sealing ring Tool One flat screwdriver No sharp edge at the head 1 One hand catches the lip of the sealing ring the other hand insert the screw...

Страница 13: ...al components must cut off power supply first And only the manufacturer or professional person designated by the manufacturer can operate 10 Transportation and Storage 10 1 The preparatory work before...

Страница 14: ...e door of the Steam Sterilizer in case of scald when open the door 9 Shut off the power before discharging installing the sealing ring and the work shall be carried out after further sufficient coolin...

Страница 15: ...13 Appendix 1 Steam loop Work Description...

Страница 16: ...14 2 Circuit diagram of sterilizer...

Страница 17: ...playing Ed B D test when put the test material onto the tray inside the chamber firstly turn on the power by the green switch close the door keeping press the UNWRAPPED button around 10s When the boar...

Отзывы: