7.5.19
Parameter P51: Overload
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
OVL MSB
Byte 1
OVL MSB
Byte 2
OVL MSB
Byte 3
OVL LSB
OVL Overload in d 0 … 9999999, default = 9 d
7.5.20
Parameter P53: A/D converter sample time (measuring rate, read only)
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
Byte 1
Byte 2
Byte 3
SAMPLETIME
SAMPLETIME
0:
6.25 Hz
==
160 ms = default
1:
2.5 Hz
==
80 ms
2:
25 Hz
==
40 ms
3:
50 Hz
==
20 ms
4:
100 Hz
==
10 ms
7.5.21
Parameter P99: Access code (Write)
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
CODE MSB
Byte 1
CODE MSB
Byte 2
CODE MSB
Byte 3
CODE LSB
CODE: If an access code has been set in the configuration tool, parameters and function
numbers for limits can be written only if the access code is also present in parameter P99.
The parameter P99 can always be written. To activate protection, P99 must be set again
to –1. If the CODE is set to 0, no access control checks will be performed.
Transmitter Series PR 5211
7 ProfiBus interface
Minebea Intec
EN-84