Byte
Description
2: STEP
One byte for the scale interval; content in decimal form: 0 to 255
1 = scale interval "1"
2 = scale interval "2"
5 = scale interval "5"
10 = scale interval "10"
20 = scale interval "20"
50 = scale interval "50"
3: LASTERROR
Last error byte; see also
CmdError
bit, number of LASTERROR:
30 = weight smaller than dead load
31 = no standstill was achieved (e.g., when taring).
33 = negative weight value when taring and W&M mode on.
35 = weight exceeds allowed range
40 = CAL switch locked
41 = transmitter not in calibration mode
42 = calibration active, transmitter is in calibration mode
46 = tare active (can occur at start calibration)
47 = no zero setting; weight not within zero setting range.
50 = invalid scale interval (step width)
51 = not enough counts/d
53 = calibration weight >FSD (Full scale deflection)
55 = arithmetic overflow
57 = entered unit does not comply with FSD weight unit
58 = SPAN above maximum
59 = fullscale cannot be divided by scale interval (stepwidth)
102, 103 = EAROM error (command SaveProcess, function number 2)
104 = wrong access code
106 = baudrate of remote display cannot be altered
107 = no standstill with
GetFixTare.
108 = parameter not valid (at entering via PLC)
7 ProfiBus interface
Transmitter Series PR 5211
EN-69
Minebea Intec
Содержание PR 5211 Series