Default: absolute.
[Standstill timeout]
If a tare- or Zero-set-command cannot be executed within time n (n = multiples of
measurement times), e. g. because the scale does not fulfill the standstill condition, the
transmitter generates a message (e. g. no standstill) and the function is aborted.
The time should be entered in multiples of measurement time, from 1 … 100, default: 8.
[Zeroset range]
Here, a +/- range is defined around the calibration zero, within which
-
the displayed gross weight can be set to zero (by an external command), or
-
the automatic zero tracking is active (see [Zerotrack repeat]).
Permitted range: 0.00 … 500.00 d, default: 50.00 d.
[Zerotrack range]
This function is only valid if the repeat interval of Zerotrack is not set to 0 (see [Zerotrack
repeat]).
Zerotrack works only as long as the weight signal is within the set zero signal.
Permitted range: 0.00 … 500.00 d, default: 0.25 d.
[Zerotrack step]
Permitted range: 0.00 … 10.00 d, default: 0.25 d.
The scale interval of the automatic Zerotrack must be smaller than the standstill range.
This function is only valid, if the repeat interval of Zerotrack is not set to 0 (see [Zerotrack
repeat]).
[Zerotrack repeat]
With the scale in standstill condition and the gross weight within the zero range, the
automatic zero tracking is done step-by-step in adjustable time intervals.
Enter multiples of the measurement time. The permitted range: is: 0 … 100, default: 0
(automatic Zerotrack = off)
-
Automatic Zerotrack can be switched off by setting Zerotrack repeat = 0.
5 Getting started
Transmitter Series PR 5211
EN-49
Minebea Intec