[Number of load cells]
Number of load cells connected in parallel (1, 2 … 10).
[Capacity of one load cell]
Maximum capacity (E
max
) of a load cell (not the total maximum capacity of the scale,
not the weighing range)
[Gravity]
Gravity at the place of installation; default is the value for Hamburg, Germany :
9.81379 m/s².
[LC 1…10 sensivity]
For this data refer to the 'Calibration Certificate' of the load cells.
In [sensivity], enter the data of the 'Rated output'.
[LC 1…10 resistance]
For this data refer to the 'Calibration Certificate' of the load cells. In [resistance], enter
the data of the 'output impedance'.
[Hysteresis correction]
Only if [with] has been selected, the values for [Correction A/B] must be entered. For
this data refer to the load cell certificate.
1.
Select [Load Cell Data].
2.
Enter the desired settings.
3.
Press [Download to PR5211/00] to save the data in PR 5211.
Transmitter Series PR 5211
5 Getting started