Field
Function
E3
The measuring signal is >36 mV (no end of conversion);
error 3
E1
Arithmetic error (overflow);
error 1
CmdError
Error during execution (CmdError); e.g., the "taring" operation is not
processed, because the scale is not at a standstill. The error is stored
in
LastError
(function number 4).
The bit is reset with the
ResetError
bit (function number 121, see
Chapter
).
CmdBusy
The device is busy executing a function (e.g., waiting for downtime
for taring).
PowerFail
Power failure; is always set after power on. The
PowerFail
bit is reset
with the
ResetPWF
bit (function number 85, see Chapter
)
"Reset power failure".
Test_Active
The device executes the ADC test. The read weight value is not the
gross value, but the test value.
Cal_Changed
The device has been calibrated.
When this bit is 1, the weighing parameters (EXPO/UNIT/STEP) must
be read again.
Set after "Power on" and reset after reading the FSD (Full scale de-
flection).
Tare_Active
The scale has been tared.
7.2.5.3
Function number 2: For internal use only.
7.2.5.4
Function number 3: For internal use only.
7.2.5.5
Function number 4: calibration information, error byte (read)
Byte
Description
0: EXPO
One byte for the position of the decimal point; content in decimal
form: 0 to 255.
0 = 000000
1 = 00000.0
2 = 0000.00
3 = 000.000
1: UNIT
One byte for the weight unit; content in decimal form: 0 to 255
1 = mg (milligrams)
2 = g (grams)
3 = kg (kilograms)
4 = t (tons)
5 = lb (pounds)
6 = l (liter)
Transmitter Series PR 5211
7 ProfiBus interface
Minebea Intec
EN-68