If the standstill timeout is exceeded in a function which needs the standstill condition, the
function will be aborted (taring parameters, set zero, calibration, 'P'-command of the
SMA-protocol). In addition, the error code 31 and the 'CmdError' bit will be set.
7.5.15
Parameter P47: Zero range
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
Byte 1
Byte 2
RANGE MSB
Byte 3
RANGE LSB
RANGE: 0 … 500.00 d, default = 50 d
Value = d x 100
7.5.16
Parameter P48: Zerotrack range
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
Byte 1
Byte 2
RANGE MSB
Byte 3
RANGE LSB
RANGE: 0 … 500.00 d, default = 0.25 d
Value = d x 100
7.5.17
Parameter P49: Zerotrack step
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
Byte 1
Byte 2
STEP MSB
Byte 3
STEP LSB
STEP: 0 … 10.00 d, default = 0.25 d
Value = d x 100
7.5.18
Parameter P50: Zerotrack repeat
Repeat interval of Zerotrack
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
Byte 1
Byte 2
Byte 3
REPEAT
REPEAT: 0 … 100 measurement times, default = 0 (Zerotrack is off)
7 ProfiBus interface
Transmitter Series PR 5211
EN-83
Minebea Intec