Error (ERR)
LED status
Description
Comments
Of
No error
No power
Constant green
Exception error
-
Module has the status "EXCEPTI-
ON"
-
The exception error monitoring
system closes all open connec-
tions to the module.
Flickering red
CRC error
1 Hz lashing red
-
Error address
-
Error baud rate
After possible changes:
-
Invalid address
-
No permissible baud rate
4.5.11
ProiNet I/O interface
The ProiNet I/O interface card has the type designation PR 1721/66 or PR 1721/76.
The ieldbus card is equipped with a standard RJ-45 socket (PR 1721/66) or two standard
RJ-45 sockets (PR 1721/76) for network connection.
It contains powerful UDP/IP connecting circuitry with transfer rates of 10 and 100 Mbit/s.
The card is inserted in the Option 1/FB slot, see Chapter
.
Technical data
Description
Data
Internal connection
Pin strip, 50
‑
pin
External connection
One RJ
‑
45 socket (PR 1721/66) or two standard RJ
‑
45 sockets
(PR 1721/76) in the module cover
Transfer rate
10 Mbit/s and 100 Mbit/s
Auto-detection (10/100, HalfDX/FullDX)
Protocol
ProiNet I/O
Connection mode
Network
Coniguration
XML ile
PR 1721/66
"GSDML-Vx.xx-Sartorius-PR5500-xxxxxx.xml"
PR 1721/76
"GSDML-Vx.xx-Minebea-PR5500-2-Port-xxxxxx.xml"
Potential isolation
Yes
Cable type
Twisted pairs, screened, e.g., patch cable CAT5 Autolink (stra-
ight or crossover)
Cable impedance
150
Ω
Cable length to HUB
Max. 115 m
Dimensions (LxWxH)
55x50x22 mm
Process Controller Maxxis 4 PR 5500
4 Device installation
EN-90