7.3.4
Supply voltage connection 230 V AC
Supply voltage
U
AC
= 100 to 240 V
+10%/-15%, 50/60 Hz
Max. power consumption
14 W/32 VA
Primary fuse
2 x 1 AT; 250 V, 5 x 20 mm;
Littlefuse series 218, order no. 0218.001.P
7.3.5
Supply voltage connection 24 V DC
Supply voltage
U
DC
= 24 V
±10%
Max. power consumption
14 W
Primary fuse
1 × 2 AT; 250 V; 5 × 20 mm;
e.g.: Schurter: SPT5 × 20, order
number: 0001.2507
7.4 Efect of ambient conditions
7.4.1
Ambient conditions
Temperature range
Ambient temperature for operation -10…+50 °C
Ambient temperature "veriiable"
-10…+40 °C
Power-on temperature
>0°C
Limits for storage/transport
-20…+70 °C
Moisture
<95%, non-condensing (acc. to IEC 60068-2)
Protection class
Front IP65, back IP20
Altitude
<2000 m
7.4.2
Electromagnetic Compatibility (EMC)
All data in compliance with NAMUR NE 21, EN 45501 and EN 61326.
High frequency electromagnetic ields
(80…3000 MHz)
EN 61000-4-3
10 V/m
Housing
Electrostatic discharge (ESD)
EN 61000-4-2
6/8 kV
Fast transients (burst)
EN 61000-4-4
1 kV
Peak voltages (surge) 1.2/50 µs
EN 61000-4-5
1/2 kV
Signal and control
lines
Conducted disturbances by high fre-
quency coupling (0.15…80 MHz)
EN 61000-4-6
10 V
7 Technical data
Process Controller Maxxis 4 PR 5500
Minebea Intec
EN-107