②
0 ext. supply
③
– ext. supply
④
Potential-free
⑤
Screen
⑥
Grounding terminal or screen clamping rail
⑦
Equipotential bonding conductor
4.5.4.7
Connecting an analog weighing platform (CAP… series)
Up to 2 Combics analog platforms (CAP… series) can be connected to the internal
weighing electronics connections depending on the model.
NOTICE
The cable colors shown here are valid, for example, for a CAPP4 500 x 400 and a
CAPP1 320 x 420.
The assignments of cable colors are listed in the relevant weighing platform
operating instructions.
Cable screens should be connected to the screen clamping rail of the device. If the
measuring lines (+M, -M) are screened, these screens must be connected to the "GNDA" in
the terminal block.
Cable screens should be connected to the grounding terminal or the screen clamping rail
of the device.
Cable colors for CAPP1 and CAPP4:
Green = green
White = white
Gray
= gray
Black = black
Red
= red
Blue
= blue
4 Device installation
Process Controller Maxxis 4 PR 5500
Minebea Intec
EN-59