Description
Data
Output temperature error
<100 ppm/K
Output zero point error
0.05%
Output max. error
<0.1%
Load
Max. 0 to 500
Ω
Potential isolation
Yes
Cables
<150 m, screened
Input
Input
1 current or voltage input
Input range
0 to 10 V or 0 to 20 mA
Input resistance
100 k
Ω
for 10 V measurement input
200
Ω
for 20 mA measurement input
Input resolution
Internal 14 bits binary = 20,000 counts,
@ e.g. 0 to 20 mA/0 to 10 V
Input max. error
0.2%
Input linearity error
<0.02%
Input temperature error
<75 ppm/K
Input thresholds
±15%, i.e., -1.5 V to +11.5 V
Potential isolation
Yes, joint inputs and outputs
Cables
<150 m, screened
Connection
Internal connection
Contact strip
External connection
2 x terminal, 6-pin
Dimensions (L×W×H)
50×45×18mm
Weight
Approx. 40 g
PR 5500/07 2x interface
o
+
o
+ o
+
A
G
ND
I
out
0…20
mA
A
G
ND
V
in
0…10
V
A
G
ND
I
in
0…20
mA
4 Device installation
Process Controller Maxxis 4 PR 5500
Minebea Intec