8
Error correction
The table lists frequent problems, as well as possible causes and corrective measures.
If the problem persists, inform Minebea Intec or an authorized dealer.
Symptom
Possible cause
Corrective measure
Switching on not
possible
Scale is not supplied wi-
th power
Check connections and voltage
Poor accuracy
Incorrect calibration
Unstable environment
Perform a calibration
Put scale in a suitable location
Application cannot
be called up
Application has not be-
en activated
Activate the application in the menu
Unit cannot be cal-
led up
Unit has not been acti-
vated
Activate the unit in the menu
Battery icon is flas-
hing
Low battery level
Connect scale to the mains and charge
battery
[Err 8.1]
Error during switching
on
Read weight exceeds start-up/zeroing
limit
[Err 8.2 ]
Error during switching
on
Read weight falls below start-up/ze-
roing limit
[Err 8.3]
Overload range error
Read weight exceeds overload limit
[Err 8.4]
Underload range error
Read weight falls below overload limit
[Err 8.5]
Tare outside of the tare
range
Adjust tare value accordingly
[Err 8.6]
Display capacity fallen
short of
Weight > 6 characters
[Err 9.5]
Incorrect calibration da-
ta
Repeat calibration
Busy
Display during tare setting, zero point
setting, printing
[--NO--]
Action not permitted
Function cannot be performed
[CAL E]
Calibration error
Unstable environment
Incorrect calibration
weight
Repeat calibration
Put scale in a suitable location
Use correct calibration weight
[REF.ERR]
Invalid reference weight The weight on the load plate is too high
or too low in order to define a valid re-
ference weight. Reduce or increase re-
ference weight
Battery cannot be
fully charged
Battery is defective
Have battery replaced by authorized
Minebea Intec service dealer.
Weight indicator Puro®
8 Error correction
EN-58
Minebea Intec