Icon
Charge status
0 to 10% remaining
11 to 40% remaining
41 to 70% remaining
71 to 100% remaining
Note:
If the battery icon flashes rapidly, then there is around 30 minutes of working time left.
When [lo.bat] is displayed, the scale switches off.
WARNING
Danger of explosion
If the rechargeable battery is replaced with a battery of the wrong type, or if it is not
connected correctly, then there is a danger of explosion.
The battery may only be replaced with the same type by an authorized Puro®
service dealer.
The battery must be disposed of according to the locally valid laws and regulations.
3.2.2
Connecting the platform
Connection of an analogue Puro® platform or a commercially available strain gauge load
cell to a Puro® indicator.
1.
Switch off the indicator and disconnect from the electrical supply.
2.
Turn over the indicator.
3.
Remove the four rubber plugs on the back.
4. Loosen the screws underneath.
5.
Remove the cover.
6. Loosen the screw connection.
7.
Run the load cell cable through the screw connection.
8. Secure the wires of the load cell cable into the terminals; when doing so, pay
attention to the pin assignments and the color coding of the connection cable for the
platform/load cell.
3 Device installation
Weight indicator Puro®
Minebea Intec
EN-9