3.
Repeat until [CAL.ADJ] is shown in the display.
4. Press the
Off
(Yes) key to access the sub-menu item.
5.
Select until [LIN] is shown in the display.
6. Press the
Off
(Yes) key to start the linearization.
[0 kg] is shown in the display. [0] flashes.
7.
Empty the load plate.
8. Press the
Off
(Yes) key to confirm that no load has been applied.
In the display, the value of the 1
st
linearization weight (50% of the calibration
weight) is displayed with flashing digits [00075.0] (in this example 75 kg). This
value cannot be changed.
[--C--] is displayed while the zero value is being saved.
9. Place the 1
st
linearization weight on the load plate.
Weight indicator Puro®
5 Operating
EN-48
Minebea Intec