If an error occurs during the linearization, [CALE] (calibration error) is displayed and
the process is canceled.
10. Press the
Off
(Yes) key to start the linearization at 50% of the calibration weight .
In the display, the value of the 2
nd
linearization weight at 100% of the calibration
weight is displayed with flashing digits [00150.0] (in this example 150 kg).
[--C--] is displayed while the value is being saved.
11. Place the 2
st
linearization weight on the load plate.
12. Press the
Off
(Yes) key to start the linearization at 100% of the calibration weight.
In the display, the value of the linearization weight at 100% of the calibration
weight is displayed [00150.0] (in this example 150 kg).
[--C--] is displayed while the value is being saved.
5 Operating
Weight indicator Puro®
Minebea Intec
EN-49