The configuration of the A/D converter is complete:
Now the [END] menu item is displayed.
5.4.2
[CAL] calibration
Note:
The calibration weight must be > 70% of the max. load, if linearization still needs to be
carried out.
Call up the menu mode:
15. Confirm the current setting with the
Off
(Yes) key or press the
Unit
key in order to
make a new setting.
If the value needs to be changed, the first figure on the left-hand side starts
flashing.
16. Use the
Unit
key or the
Mode
F
key to select a new digit and confirm this with the
Off
key.
The next digit starts flashing.
17. Repeat the process in order to enter a new calibration weight.
18. Confirm the new value with the
Off
key.
19. Press the
Menu
M
+
key to exit the setup menu.
In the next step the scale is calibrated.
The configuration of the A/D converter is complete.
1.
Press and hold
Menu
M
+
until [M.E.N.U] is displayed.