— ON
Locked
— S.CAL.AD
Calibration/adjustment menu
— OFF
Unlock*
— ON
Locked
— RESET
Restore factory setting of the current menu
— OFF
Unlock*
— ON
Locked
— END
Exit menu level
— CAL
Initiates a two-point calibration (zero and max.
load)
— LIN
Initiates a linearization (zero, half weighing
range and max. load)
— GEO
The adjustment of the calibration based on the
current location is carried out using the
geographic adjustment factor (GEO). (Settings
from 0 … 31, default value = 12)
— END
Exit menu level
5.4
Calibration and adjustment
The scale can be calibrated and adjusted in this menu.
Initial calibration
If the scale is being put into operation for the first time, calibration is recommended in
order to ensure precise weighing results. Before the calibration, ensure that the
appropriate calibration weights are available.
Adjust the GEO setting according to the location (see Table
).
CAL.ADJ
5.4.1
[ADC.CON] configuring A/D converter
Note:
If the calibration weight has been changed, the linearization weights also change. See
Chapter
Configuring the A/D converter:
1.
Press and hold the
Menu
M
+
(Menu) key.
The [M.E.N.U] menu item is displayed briefly, then the display switches
automatically to the [APPLIC] menu item.