5.6.7
Set preload
Setup information
-
Setting a preload for scales used in legal metrology is automatically carried out by the
power-on zero range. Explicit setting of the preload is not possible in legal
metrology.
After approx. 2 seconds, the prompt appears to place the irst linearization
weight on the scale.
5. Place the required weight on the scale.
After a short time, the diference between the measured value and the true
weight is displayed.
6. Accept the diference by pressing
(cancel by pressing
).
The prompt appears to place the second linearization weight on the scale.
7. Repeat the procedure for all required linearization weights.
After the inal linearization weight has been stored, a prompt appears asking you
to unload the load plate completely.
8. Remove the load from the load plate.
After a short time, the zero point is automatically adopted and the automatically
switches to weighing mode.
9. Close the locking switch again and, if necessary, reconigure the key
(press and
hold) to calibration/adjustment beforehand.
5 Getting started
Combics 1 Series Complete Scales Models CAW1P, CAW1S, CAS1
Minebea Intec
EN-55
Содержание CAS1
Страница 18: ...CAS1G IG Combics 1 Series Complete Scales Models CAW1P CAW1S CAS1 3 Device description EN 16 Minebea Intec ...
Страница 128: ......
Страница 136: ...Typ Waage Minecomb Typ Auswertegerät TA EG Bauartzulassung T11379 Prüfschein D09 11 02 Typ Minecomb ...
Страница 137: ...Typ Waage Minecomb Typ Auswertegerät TA EG Bauartzulassung T11379 Prüfschein D09 11 02 ...
Страница 138: ...Option Y2 Typ Waage Minecomb Typ Auswertegerät TA EG Bauartzulassung T11379 Prüfschein D09 11 02 ...
Страница 140: ...CAW3S1 60FE NCE Typ Waage Minecomb Typ Auswertegerät TA EG Bauartzulassung T11379 Prüfschein D09 11 02 ...
Страница 141: ......
Страница 142: ......
Страница 143: ...1 5 ...
Страница 150: ......
Страница 151: ......
Страница 152: ......
Страница 153: ......
Страница 154: ......
Страница 155: ......