Electrical Safety Inspection B-5
normal polarity( Normal Condition);
reverse polarity( Normal Condition),
normal polarity with open neutral(Single Fault Condition);
reverse polarity with open neutral(Single Fault Condition).
normal polarity with open earth(Single Fault Condition);
reverse polarity with open earth(Single Fault Condition).
LIMITS
For CF
applied parts,
10
μ
A in Normal Condition
50
μ
A in Single Fault Condition
For BF
applied parts,
100
μ
A in Normal Condition
500
μ
A in Single Fault Condition
NOTE:
Make sure the safety analyzer is authorized comply with requirement of IEC60601-1.
Follow the instructions of the analyzer manufacturer.